Comparing the dosimetric characteristics of the electron beam from dedicated intraoperative and conventional radiotherapy accelerators

نویسندگان

  • Hamid Reza Baghani
  • Seyed Mahmoud Reza Aghamiri
  • Seyed Rabi Mahdavi
  • Mohammad Esmail Akbari
  • Hamid Reza Mirzaei
چکیده

The specific design of the mobile dedicated intraoperative radiotherapy (IORT) accelerators and different electron beam collimation system can change the dosimetric characteristics of electron beam with respect to the conventional accelerators. The aim of this study is to measure and compare the dosimetric characteristics of electron beam produced by intraoperative and conventional radiotherapy accelerators. To this end, percentage depth dose along clinical axis (PDD), transverse dose profile (TDP), and output factor of LIAC IORT and Varian 2100C/D conventional radiotherapy accelerators were measured and compared. TDPs were recorded at depth of maximum dose. The results of this work showed that depths of maximum dose, R90, R50, and RP for LIAC beam are lower than those of Varian beam. Furthermore, for all energies, surface doses related to the LIAC beam are substantially higher than those of Varian beam. The symmetry and flatness of LIAC beam profiles are more desirable compared to the Varian ones. Contrary to Varian accelerator, output factor of LIAC beam substantially increases with a decrease in the size of the applicator. Dosimetric characteristics of beveled IORT applicators along clinical axis were different from those of the flat ones. From these results, it can be concluded that dosimetric characteristics of intraoperative electron beam are substantially different from those of conventional clinical electron beam. The dosimetric characteristics of the LIAC electron beam make it a useful tool for intraoperative radiotherapy purposes.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Commissioning the First Mobile Dedicated Accelerator for Intraoperative Electron Radiotherapy in Iran

Introduction: Intraoperative radiotherapy is a radiotherapy technique in which a high single fraction of radiation dose is delivered to the patient after surgery and Concurrent with anesthesia time. The most frequent method for IORT implementation is Intraoperative electron radiotherapy (IOERT), in which, some dedicated and high dose per pulse electron accelerators are employe...

متن کامل

Physical characteristics of electron beam from conventional and beam shaper IOERT applicator: A comparison study

Introduction: Intraoperative electron radiation therapy (IOERT) is one of the cancer treatment techniques that delivers high doses to tumor bed during surgery. IOERT can be performed by either conventional LINACs or dedicated IORT accelerators such as LIAC (Light Intraoperative Accelerator). Two types of applicators can be used with LIAC dedicated accelerator including conventi...

متن کامل

اعتبارسنجی سر شتاب‌دهندۀ پرتودرمانی حین عمل‌ جراحی LIAC با استفاده از روش مونت‌کارلو

With the development of science and technology in the medical field and need more than ever to accuracy and quality of treatment has been causer increase different radiotherapy methods. Almost the past two decade’s intraoperative radiation therapy (IORT) as one of the new techniques used to treat cancer patients. One of the problems with this method is obtaining accurate dosimetry, becaus...

متن کامل

Monte Carlo Simulation of Electron Beams produced by LIAC Intraoperative Radiation Therapy Accelerator

Background: One of the main problems of dedicated IORT accelerators is to determine dosimetric characteristics of the electron beams. Monte Carlo simulation of IORT accelerator head and produced beam will be useful to improve the accuracy of beam dosimetry.Materials and Methods: Liac accelerator head was modeled using the BEAMnrcMonte Carlo simulation system. Phase-space files were generated at...

متن کامل

شبیه‌سازی شتاب‌ دهنده پرتو درمانی حین عمل LIAC به همراه اپلیکاتور شکل دهنده باریکه بااستفاده از محاسبات مونت‌کارلو

Intraoperative electron radiotherapy is one of the radiotherapy methods that delivers a high single fraction of radiation dose to the patient in one session during the surgery. Beam shaper applicator is one of the applicators that is recently employed with this radiotherapy method. This applicator has a considerable application in treatment of large tumors. In this study, the dosimetric charact...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 16  شماره 

صفحات  -

تاریخ انتشار 2015