Evaluation of Several Dosimeters for Identification of Irradiated Foods Using a 5 MeV Electron Beam

نویسندگان

  • Makoto Miyahara
  • Hitoshi Ito
  • Kouji Ueno
  • Yutaka Yamase
  • Masatake Toyoda
چکیده

Electron-beam irradiation facilities are commercially utilized for food irradiation. This study investigates whether dosimeters used to detect gamma-rays can also be used to accurately detect electron-beam irradiation. The irradiation souses for gamma-ray at the Takasaki Institute were used throughout this study. Four kinds of dosimeters, RadiaChromic (RC), GammaChromic (GC), Amber3042 (AM), Radix (RX), and alanine (AL) were irradiated at doses from 0.1 to 60 kGy using a 5 MeV electron-beam (EB) at the Japan Irradiation Service Co. Ltd. (Japan). AL was submitted to the National Physics Laboratory in order to evaluate the doses that AL were absorbed. At low-level dose range, all dosimeters indicated correct values. This means those dosimeters were usable to accurately detect electron beam irradiation. But at higher range they all required correction due to either temperature effects or varying sensitivity to electron-beam exposure. Dose–depth profiles were obtained using ham and cheese samples, in order to check penetration of EB by the calibrated machine.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Verification of dose rate and energy dependence of MAGICA polymer gel dosimeter with electron beams

Background: The purpose of this study was to evaluate the dependency of MAGICA polymer gel dosimeter response (R2) on different electron energies as well as on different mean dose rate for a standard clinically used linear accelerator. Materials and Methods: The sensitivity of the dosimeter was represented by the slope of calibration curve in the linear region measured for each modalit...

متن کامل

Electron Beam Dosimetry in Heterogeneous Phantoms Using a MAGIC Normoxic Polymer Gel

Introduction: Nowadays radiosensitive polymer gels are used as a reliable dosimetry tool for verification of 3D dose distributions. Special characteristics of these dosimeters have made them useful for verification of complex dose distributions in clinical situations. The aim of this work was to evaluate the capability of a normoxic polymer gel to determine electron dose distributions in differ...

متن کامل

Evaluation of LiF:Mg,Ti (TLD-100) for Intraoperative Electron Radiation Therapy Quality Assurance

BACKGROUND Purpose of the present work was to investigate thermoluminescent dosimeters (TLDs) response to intraoperative electron radiation therapy (IOERT) beams. In an IOERT treatment, a large single radiation dose is delivered with a high dose-per-pulse electron beam (2-12 cGy/pulse) during surgery. To verify and to record the delivered dose, in vivo dosimetry is a mandatory procedure for qua...

متن کامل

Investigating the Effects of Cut-Out Shield on High-Energy Electron Fields Using MAGIC Normoxic Polymer Gel

Introduction The use of cut-outs in electron applicators make changes on output, isodose, and percentage depth dose (PDD) curves. These changes and electron beam dose distribution in the form of three-dimensional (3D) can be measured by gel dosimeters. Materials and Methods Dosimetry was performed with and without a square shield (6×6 cm2 field). The energies were 4, 9, and 16 MeV and phantom w...

متن کامل

Evaluation of Dose Calculation Accuracy of Isogray Treatment Planning System in Craniospinal Radiotherapy

Introduction: Craniospinal radiotherapy is a therapeutic technique for central nervous system (CNS) tumors, which requires meticulous attention to technique and dosimetry.Treatment planning system (TPS) is one of the main equipment in radiotherapy; therefore, the evaluation of its accuracy is essential for dose calculation. The present study evaluates the validity of Isogray TPS in craniospinal...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2002