First-in-man use of a novel embolic protection device for patients undergoing transcatheter aortic valve implantation.
نویسندگان
چکیده
AIMS We describe the first-in-human experience with a novel cerebral embolic protection device used during transcatheter aortic valve implantation (TAVI). One current challenge of TAVI is the reduction of procedural stroke. Procedural mobilisation of debris is a known source of cerebral embolisation. Mechanical protection by transient filtration of cerebral blood flow might reduce the embolic burden during TAVI. We aimed to evaluate the feasibility and safety of the Claret CE Pro™ cerebral protection device in patients undergoing TAVI. METHODS AND RESULTS Patients scheduled for TAVI were prospectively enrolled at three centres. The Claret CE Pro™ (Claret Medical, Inc. Santa Rosa, CA, USA) cerebral protection device was placed via the right radial/brachial artery prior to TAVI and was removed after the procedure. The primary endpoint was technical success rate. Secondary endpoints encompassed procedural and 30-day stroke rates, as well as device-related complications. Deployment of the Claret CE Pro™ cerebral protection device was intended for use in 40 patients, 35 devices were implanted into the aortic arch. Technical success rate with delivery of the proximal and distal filter was 60% for the first generation device and 87% for the second-generation device. Delivery times for the first-generation device were 12.4±12.1 minutes and 4.4 ± 2.5 minutes for the second-generation device (p<0.05). The quantity of contrast used related to the Claret CE Pro System was 19.6 ± 3.8 ml. Captured debris was documented in at least 19 of 35 implanted devices (54.3%). No procedural transient ischaemic attacks, minor strokes or major strokes occurred. Thirty-day follow-up showed one minor stroke occurring 30 days after the procedure, and two major strokes both occurring well after the patient had completed TAVI. CONCLUSIONS The use of the Claret CE Pro™ system is feasible and safe. Capture of debris in more than half of the patients provides evidence for the potential to reduce the procedural cerebral embolic burden utilising this dedicated filter system during TAVI.
منابع مشابه
Cerebral Embolic Protection Devices During Transcatheter Aortic Valve Implantation: Systematic Review and Meta-Analysis.
BACKGROUND AND PURPOSE Silent ischemic embolic lesions are common after transcatheter aortic valve implantation (TAVI). The use of embolic protection devices (EPD) may reduce the occurrence of these embolic lesions. Thus, a quantitative overview and credibility assessment of the literature was necessary to draw a robust message about EPD. Therefore, the aim of this meta-analysis was to study wh...
متن کاملTranscatheter Aortic Valve Replacement: A Review Article
Transcatheter aortic valve replacement (TAVR) is a novel therapeutic intervention for the replacement of severely stenotic aortic valves in high-risk patients for standard surgical procedures. Since the initial PARTNER trial results, use of TAVR has been on the rise each year. New delivery methods and different valves have been developed and modified in order to promote the minimally invasive p...
متن کاملIncidence and Prevention of Strokes in TAVI
Transcatheter aortic valve implantation (TAVI) is now a widely adopted option for many inoperable and high risk patients with severe aortic valve stenosis, and clinical trials continue to show great benefit with regards to mortality and major cardiovascular endpoints. As the technology continues to expand and possibly grow to include intermediate and low risk populations, investigators have rem...
متن کاملLetter by Fanning et al regarding article, "Histopathology of embolic debris captured during transcatheter aortic valve replacement".
We congratulate Van Mieghem and colleagues for the significant contribution they have made to the understanding of the embolic phenomena associated with transcatheter aortic valve implantation through their article, " Histopathology of Embolic Debris Captured During Transcatheter Aortic Valve Replacement. " 1 Somewhat surprising is the high occurrence of thrombotic material, which was identifie...
متن کاملRetinal embolic events: frequency and impact following transcatheter aortic valve implantation (TAVI) for aortic stenosis
Objectives Transcatheter aortic valve implantation (TAVI) is an established treatment for patients with severe symptomatic aortic stenosis. It has a cerebrovascular accident rate of about 5% but the effect on retinal embolic events has not been previously reported. This study investigated the occurrence of retinal emboli following TAVI. Methods and analysis In this prospective observational s...