The clinical significance of vacuum mixing bone cement.

نویسندگان

  • M H Geiger
  • E M Keating
  • M A Ritter
  • J A Ginther
  • P M Faris
  • J B Meding
چکیده

This controlled study compared the strength and porosity of 48 polymethylmethacrylate cement-implant constructs prepared with open bowl versus vacuum mix technique. Forty-eight blast finished stainless steel rods of 13 mm diameter were implanted with centralizers into 17-mm inner diameter tubes that had been retrograde filled with polymethylmethacrylate cement. The eight cement preparations used were open bowl and vacuum mixed Simplex, Osteobond, Zimmer Dough Type, or Palacos R. Six replications of each condition were performed. The tubes were maintained at 37 degrees C. Each tube was cut transversely into five segments. The center three segments were used for data analysis: pushout strength, cycles to failure, and interface porosity analysis. Rod pushout data showed there was no significant difference between open bowl and vacuum mixed samples when all cement brands were combined. Mean sheer force for Palacos R vacuum mixed samples was greater than open bowl (634+/-47 versus 423+/-171), whereas the force for the Zimmer Dough Type cement open bowl was greater than that of the vacuum mixed samples (901+/-71 versus 705 +/-82). Cycles to failure data did not show significant differences when open bowl and vacuum mixed samples were compared when cements were analyzed individually or combined. Image analysis of cement-implant interfaces showed that vacuum mixing reduced void area significantly compared with open bowl mixing in the Palacos R and Osteobond preparations. Vacuum mixing does not appear to reduce cement prosthesis interface porosity or improve its mechanical properties in all cements.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Acrylic antibiotic-loaded bone cement: a basic study

  Abstract   Objective: The aim of this study was to evaluate the efficacy of antibiotic-loaded bone   cement in controlling local infection and in regard to its physical characteristics, elastic   modulus, and tensile strength in-vitro.   Methods: Acrylic bone cement, based on polymethylmethacrylate (PMMA) was   mixed with the powder form of three antibiotics, i.e., gentamicin, tobramycin, and...

متن کامل

Vacuum-mixing cement does not decrease overall porosity in cemented femoral stems: an in vitro laboratory investigation.

The role of vacuum mixing on the reduction of porosity and on the clinical performance of cemented total hip replacements remains uncertain. We have used paired femoral constructs prepared with either hand-mixed or vacuum-mixed cement in a cadaver model which simulated intra-operative conditions during cementing of the femoral component. After the cement had cured, the distribution of its poros...

متن کامل

The effect of vacuum mixing and pre-heating the femoral component on the mechanical properties of the cement mantle.

We investigated the effect of pre-heating a femoral component on the porosity and strength of bone cement, with or without vacuum mixing used for total hip replacement. Cement mantles were moulded in a manner simulating clinical practice for cemented hip replacement. During polymerisation, the temperature was monitored. Specimens of cement extracted from the mantles underwent bending or fatigue...

متن کامل

Comparison of various vacuum mixing systems and bone cements as regards reliability, porosity and bending strength.

BACKGROUND There are several vacuum mixing systems on the market which are arbitrarily used with various bone cements in clinical work. Hardly any studies have been done on the performance and handling of these systems in combination with different cement brands. MATERIAL AND METHODS We therefore tested 6 vacuum mixing systems (Palamix, Summit, Cemvac, Optivac, Vacumix, MixOR) in combination ...

متن کامل

Mixing of Acrylic Bone Cement – Cu

Dunne, NJ, Daly, C, Beverland, D Queens University of Belfast, Belf Musgrave Park Hospital, Belfast Introduction It has been shown that acrylic bone cement is weakened by its porosity, which enhances the formation of microcracks that contribute to major crack propagation . It has also been observed, that mixing procedures play a significant role in determining the quality of bone cement produce...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical orthopaedics and related research

دوره 382  شماره 

صفحات  -