Peri-implant assessment via cone beam computed tomography and digital periapical radiography: an ex vivo study
نویسندگان
چکیده
OBJECTIVES This research evaluated detail registration in peri-implant bone using two different cone beam computer tomography systems and a digital periapical radiograph. METHODS Three different image acquisition protocols were established for each cone beam computer tomography apparatus, and three clinical situations were simulated in an ex vivo fresh pig mandible: buccal bone defect, peri-implant bone defect, and bone contact. Data were subjected to two analyses: quantitative and qualitative. The quantitative analyses involved a comparison of real specimen measures using a digital caliper in three regions of the preserved buccal bone - A, B and E (control group) - to cone beam computer tomography images obtained with different protocols (kp1, kp2, kp3, ip1, ip2, and ip3). In the qualitative analyses, the ability to register peri-implant details via tomography and digital periapical radiography was verified, as indicated by twelve evaluators. Data were analyzed with ANOVA and Tukey's test (α=0.05). RESULTS The quantitative assessment showed means statistically equal to those of the control group under the following conditions: buccal bone defect B and E with kp1 and ip1, peri-implant bone defect E with kp2 and kp3, and bone contact A with kp1, kp2, kp3, and ip2. Qualitatively, only bone contacts were significantly different among the assessments, and the p3 results differed from the p1 and p2 results. The other results were statistically equivalent. CONCLUSIONS The registration of peri-implant details was influenced by the image acquisition protocol, although metal artifacts were produced in all situations. The evaluators preferred the Kodak 9000 3D cone beam computer tomography in most cases. The evaluators identified buccal bone defects better with cone beam computer tomography and identified peri-implant bone defects better with digital periapical radiography.
منابع مشابه
Comparison of Diagnostic Accuracy of Cone Beam Computed Tomography and Digital Radiography for Detection of Vertical Root Fractures with and without Gutta Percha
Objective: Diagnosis of vertical root fractures (VRFs) is critical in endodontics which most of the times occurs in endodontically treated teeth with root canal fillings such as gutta percha. Despite Cone Beam Computed Tomography (CBCT) has significantly enhanced image quality compared to digital radiography (DR) which aid the diagnosis, artifacts has remained as a problem in VRF detection . ...
متن کاملComparing the Accuracy of Cone Beam Computed Tomography,Digital Intraoral Radiography and Conventional Intraoral Radiography in the Measurement of Periodontal Bone Defects
Background and Aim: Cone beam computed tomography (CBCT) produces high-quality data in periodontal diagnosis and treatment planning. The aim of this study was to compare the accuracy of CBCT with intraoral digital and conventional radiography in the measurement of periodontal bone defects. Methods and Materials: In this diagnostic research, two hundred and eighteen artificial osseous d...
متن کاملDiagnostic Accuracy of CBCT with Different Voxel Sizes and Intraoral Digital Radiography for Detection of Periapical Bone Lesions: An Ex-Vivo Study
OBJECTIVES This study sought to assess the diagnostic accuracy of cone beam computed tomography (CBCT) with different voxel sizes and intraoral digital radiography with photostimulable phosphor (PSP) plate for detection of periapical (PA) bone lesions. MATERIALS AND METHODS In this ex vivo diagnostic study, one-millimeter defects were created in the alveolar sockets of 15 bone blocks, each wi...
متن کاملA Comparative Study of Diagnostic Accuracy of Cone Beam Computed Tomography and Intra Oral Digital Radiography Phosphor Storage Plates in Detecting Strip Perforation in Endodontically-Treated Teeth
Background and purpose: Perforation impairs the prognosis of endodontic treatments, so, timely and accurate diagnosis is important. For this purpose, various radiographic diagnostic tools have been studied to select the method that has the least complications and helps in making the most accurate diagnosis. This study aimed to compare the accuracy of cone beam computed tomography (CBCT) and int...
متن کاملComparison of cone beam computed tomography and digital radiography in detection of pulpal exposure in external cervical root resorption
Background and Aims: Pulpal exposures originated from the external cervical root resorptions have major effects on the treatment and prognosis. Therefore, the aim of this study was performed to compare the two different imaging systems-digital radiography with PSP (Photostimulable Phosphor) sensor in three horizontal different views and CBCT (cone beam computed tomography) images to assess the ...
متن کامل