Focused tight dressing does not prevent cochlear implant magnet migration under 1.5 Tesla MRI
نویسندگان
چکیده
We report a retrospective case of inner magnet migration, which occurred after 1.5 Tesla MRI scanning in an adult recipient of a bilateral cochlear implant (CI) despite a focused head dressing. The patient, bilaterally implanted with Nucleus 5 CIs (Cochlear LTD, Sydney, Australia), underwent a 1.5 Tesla cholangio-MRI scan for biliary duct pathology. In subsequent days, a focal skin alteration appeared over the left inner coil. Plain skull radiographs showed partial magnet migration on the left side. Surgical exploration confirmed magnet twisting; the magnet was effectively repositioned. Left CI performance was restored to pre-migration level. The wound healed without complications. Thus, focused dressing does not prevent magnet migration in CI recipients undergoing 1.5 Tesla MRI. All patients should be counselled on this potential complication. A minor surgical procedure is required to reposition the magnet. Nevertheless, timely diagnosis is necessary to prevent skin breakdown and subsequent device contamination. Plain skull radiograph is very effective in identifying magnet twisting; it should be performed systematically after MRI or minimally on all suspected cases.
منابع مشابه
Magnet Dislocation: An Increasing and Serious Complication Following MRI in Patients with Cochlear Implants Magnetdislokation – eine zunehmende und folgenreiche Komplikation nach MRT bei Patienten mit Cochlea Implantat
Purpose: Cochlear implantation (CI) represents the gold standard in the treatment of children born deaf and postlingually deafened adults. Initial magnetic resonance imaging (MRI) was contraindicated in CI users. Meanwhile, there are specific recommendations concerning MRI compatibility depending on the type of CI system and the device manufacturer. Some CI systems are even approved for MRI wit...
متن کاملMagnet removal and reinsertion in a cochlear implant recipient undergoing brain MRI.
BACKGROUND Recent advances in cochlear implants (CIs) and in magnetic resonance imaging (MRI) techniques have led to increasing numbers of implanted patients who have undergone brain MRIs for various reasons. This paper first describes the surgical procedure for removing the magnet before an MRI and its subsequent reinsertion immediately afterwards in a CI recipient. METHOD After the administ...
متن کاملMagnetic resonance imaging and cochlear implants: compatibility and safety aspects.
With cochlear implants, magnetic resonance imaging (MRI) has until recently been contraindicated due to excessive magnetic and electromagnetic interference. The aim of this study was to determine the MRI compatibility of the Med-E1 Combi 40/40+ cochlear implant, within a wide range of clinical MRI applications. In vitro experiments on a 1.5 T MR scanner were performed. Torque, force, demagnetiz...
متن کاملCochlear implants and 1.5 T MRI scans: the effect of diametrically bipolar magnets and screw fixation on pain
BACKGROUND The probability that a patient will need an MRI scan at least once in a lifetime is high. However, MRI scanning in cochlear implantees is associated with side effects. Moreover, MRI scan-related artifacts, dislodging magnets, and pain are often the most frequent complications. The aim of this study was to evaluate the occurrence of pain in patients with cochlear implant systems using...
متن کاملMagnetic resonance imaging investigation of the bone conduction implant – a pilot study at 1.5 Tesla
PURPOSE The objective of this pilot study was to investigate if an active bone conduction implant (BCI) used in an ongoing clinical study withstands magnetic resonance imaging (MRI) of 1.5 Tesla. In particular, the MRI effects on maximum power output (MPO), total harmonic distortion (THD), and demagnetization were investigated. Implant activation and image artifacts were also evaluated. METHO...