Dose accuracy and injection force of different insulin glargine pens.
نویسندگان
چکیده
BACKGROUND Dose accuracy and injection force, representing key parameters of insulin pens, were determined for three pens delivering insulin glargine-based copies, Pen Royale (WR) and DispoPen (WD) for Glaritus® (Wockhardt) and GanLee Pen (GL) for Basalin® (Gan & Lee), compared with pens of the originator, ClikSTAR® (CS) and SoloSTAR® (SS) for Lantus® (Sanofi). METHODS Using the weighing procedure recommended by DIN EN ISO 11608-1:2000, dose accuracy was evaluated based on nonrandomized delivery of low (5 U), mid (30 U), and high (60 U) dosage levels. Injection force was measured by dispensing the maximum dose of insulin (60 U for the GL, WR, and WD; 80 U for the SS and CS) at dose speeds of 6 and 10 U/s. RESULTS All tested pens delivered comparable average doses within the DIN EN ISO 11608-1:2000 limits at all dosage levels. The GL revealed a higher coefficient of variation (CV) at 5 U, and the WR and WD had higher CVs at all dosage levels compared with the CS and SS. Injection force was higher for the WR, WD, and GL compared with the CS and SS at both dose speeds. In contrast to the CS and SS with an end-of-content feature, doses exceeding the remaining insulin could be dialed with the WR, GL, and WD and, apparently, dispensed with the WD. CONCLUSIONS All pens fulfilled the dose accuracy requirements defined by DIN EN ISO 11608-1:2000 standards at all three dosage levels, with the WR, WD, and GL showing higher dosage variability and injection force compared with the SS and CS. Thus, the devices that deliver insulin glargine copies show different performance characteristics compared with the originator.
منابع مشابه
Accuracy and Injection Force of the Gla-300 Injection Device Compared With Other Commercialized Disposable Insulin Pens
BACKGROUND To deliver insulin glargine 300 U/mL (Gla-300), the widely used SoloSTAR(®) pen has been modified to allow for accurate and precise delivery of required insulin units in one-third of the volume compared with insulin glargine 100 U/mL, while improving usability. Here we compare the accuracy and injection force of 3 disposable insulin pens: Gla-300 SoloSTAR(®), FlexPen(®), and KwikPen™...
متن کاملInjection force of SoloSTAR® compared with other disposable insulin pen devices at constant volume flow rates.
BACKGROUND Injection force is a particularly important practical aspect of therapy for patients with diabetes, especially those who have dexterity problems. This laboratory-based study compared the injection force of the SoloSTAR® insulin pen (SoloSTAR; sanofi-aventis) versus other available disposable pens at injection speeds based on the delivered volume of insulin released at the needle. M...
متن کاملDose accuracy and injection force of disposable pens delivering pramlintide for the treatment of diabetes.
BACKGROUND The pen injection format, typically used for insulin administration, has been adapted for the injectable, noninsulin diabetes therapy pramlintide. Administered before major meals, pramlintide therapy requires two to four injections/day in addition to the patients' usual insulin injections. The dose accuracy and injection force was determined for the 60 and 120 µg pramlintide pens. ...
متن کاملTreatment intensification with insulin glargine in patients with inadequately controlled type 2 diabetes improves glycaemic control with a high treatment satisfaction and no weight gain.
PRINCIPLES We aimed to evaluate the efficacy of, and treatment satisfaction with, insulin glargine administered with SoloSTAR® or ClikSTAR® pens in patients with type 2 diabetes mellitus managed by primary care physicians in Switzerland. METHODS A total of 327 patients with inadequately controlled type 2 diabetes were enrolled by 72 physicians in this prospective observational study, which ai...
متن کاملAnalysis: desirable attributes of insulin injection pens that drive patient preference and compliance.
Insulin pens are used by approximately half of worldwide insulin users. Insulin pens have made insulin injections easier compared to traditional vial and syringe injections. In an article in this issue of Journal of Diabetes Science and Technology, Dr. Asakura discusses several important design parameters, which are considered during refillable insulin-injection pen design. Ease of cartridge re...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of diabetes science and technology
دوره 7 5 شماره
صفحات -
تاریخ انتشار 2013