investigation of collimator influential parameter on spect image quality: a monte carlo study
نویسندگان
چکیده
background: obtaining high quality images in single photon emission tomography (spect) device is the most important goal in nuclear medicine. because if image quality is low, the possibility of making a mistake in diagnosing and treating the patient will rise. studying effective factors in spatial resolution of imaging systems is thus deemed to be vital. one of the most important factors in spect imaging in nuclear medicine is the use of an appropriate collimator for a certain radiopharmaceutical feature in order to create the best image as it can be effective in the quantity of full width at half maximum (fwhm) which is the main parameter in spatial resolution. method: in this research, the simulation of the detector and collimator of spect imaging device, model hd3 made by philips co. and the investigation of important factors on the collimator were carried out using mcnp-4c code. results: the results of the experimental measurments and simulation calculations revealed a relative difference of less than 5% leading to the confirmation of the accuracy of conducted simulation mcnp code calculation. conclusion: this is the first essential step in the design and modelling of new collimators used for creating high quality images in nuclear medicine.
منابع مشابه
Investigation of Collimator Influential Parameter on SPECT Image Quality: a Monte Carlo Study
Background: Obtaining high quality images in Single Photon Emission Tomography (SPECT) device  is the most important goal in nuclear medicine. Because if image quality is low, the possibility of making a mistake in diagnosing and treating the patient will rise. Studying effective factors in spatial resolution of imaging systems is thus deemed to be vital. One of the most important factors in S...
متن کاملInvestigation of Collimator Influential Parameter on SPECT Image Quality: a Monte Carlo Study
BACKGROUND Obtaining high quality images in Single Photon Emission Tomography (SPECT) device is the most important goal in nuclear medicine. Because if image quality is low, the possibility of making a mistake in diagnosing and treating the patient will rise. Studying effective factors in spatial resolution of imaging systems is thus deemed to be vital. One of the most important factors in SPE...
متن کاملEvaluation of the performance of parallel-hole collimator for high resolution small animal SPECT: A Monte Carlo study
Introduction: Image quality and accuracy of in vivo activity quantification in SPECT are affected by collimator penetration and scatter components, especially in high energy imaging. These phenomena highly depend on the collimator characteristic and photon energy. The presence of penetrated and scattered photons from collimator in SPECT images degrades spatial resolution, contr...
متن کاملAttenuation Correction in SPECT during Image Reconstruction using an Inverse Monte Carlo Method: A Simulation Study
Introduction: The main goal of SPECT imaging is to determine activity distribution inside the organs of the body. However, due to photon attenuation, it is almost impossible to do a quantitative study. In this paper, we suggest a mathematical relationship between activity distribution and its corresponding projections using a transfer matrix. Monte Carlo simulation was used to find a precise tr...
متن کاملSPECT Uptake Ratios: A Monte Carlo Study
In dopamine brain SPECT, semiquantitative techniques are in use, mostly for research purposes, to calculate activity uptake in the striatum relative to the background. The measured uptake ratios depend on both acquisition and reconstruction, and one important parameter is the rotation radius of the g-camera detectors, which affects spatial resolution. In brain SPECT research studies, the rotati...
متن کاملA Monte Carlo study of SPECT in boron neutron capture therapy for a heterogeneous human phantom
Background: Boron neutron capture therapy (BNCT) is a binary radiotherapy combining biochemical targeting with neutron irradiation. However, monitoring the boron distribution is a fundamental problem in BNCT. Prompt gamma rays emitted by boron capture reaction can be used to address the issue. Materials and Methods: The general-purpose Monte Carlo toolkits Geant4 and MCNP were used for the simu...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
journal of biomedical physics and engineeringجلد ۵، شماره ۱ Mar، صفحات ۰-۰
کلمات کلیدی
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023