quantitative evaluation of lif: mg, ti (tld-100) long-term sensitivity stability
نویسندگان
چکیده
background : sensitivity is an important dosimetric issue particularly in personal dosimetry where the range of doses is often very low to medium. in-service aging is assumed to be of negligible effect on the response of thermoluminescent dosimeters (tld), consequently, some service providers may mix fresh tlds to in-use ones and simply normalize the readings of tlds through batch calibration. this may lead to an increase in the coefficient of variation of the batch. objective: the correlation between the sensitivity of lif(mg:ti) and its in-service life is quantitatively studied in this paper. method: three groups of dosimeters with different service lives were chosen from the inventory of the national tld service provider. the tlds were then exposed to different doses and their tl responses were measured and compared. result: by analyzing the results obtained from groups with different in-service ages, it was shown that the sensitivity of the dosimeters degrades with age in long term, and the amount of this sensitivity loss was measured. the reading standard deviation of the dosimeters degrades with the service life for low doses but seems to be insensitive to the service life in medium and high doses. conclusion: a more efficient annealing approach is recommended time to time for tlds. moreover, the calibration dosimeters must be updated whenever a big difference between the sensitivity of new and used tlds is seen.
منابع مشابه
Investigation of LiF, Mg and Ti (TLD-100) Reproducibility
LiF, Mg and Ti cubical TLD chips (known as TLD-100) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mGy) with 662keV photons of Cs-137. A group of 40 cubical TLD chips was randomly selected from a batch and the values of Element Correction Coefficient (ECC) were obtained 4 times by irradiating them to doses...
متن کاملInvestigation of LiF, Mg and Ti (TLD-100) Reproducibility
LiF, Mg and Ti cubical TLD chips (known as TLD-100) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mGy) with 662keV photons of Cs-137. A group of 40 cubical TLD chips was randomly selected from a batch and the values of Element Correction Coefficient (ECC) were obtained 4 times by irradiating them to doses...
متن کاملinvestigation of lif, mg and ti (tld-100) reproducibility
lif, mg and ti cubical tld chips (known as tld-100) are widely used for dosimetry purposes. the repeatability of tl dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mgy) with 662kev photons of cs-137. a group of 40 cubical tld chips was randomly selected from a batch and the values of element correction coefficient (ecc) were obtained 4 times by irradiating them to doses...
متن کاملEvaluation of the Entrance Skin Dose in Animals Undergoing Diagnostic Radiology Using LiF, Mg, Ti Thermoluminescence Dosimetry (TLD-100)
Introduction According to the International Commission on Radiological Protection publication numbers 60 and 21, the environmental control standards should ensure human and other species protection to a desirable degree. Since application of radiographic procedures in Veterinary Medicine has increased significantly, in this study, we aimed to evaluate the entrance skin dose to the animals (e.g....
متن کاملتولید دزیمتر LiF: Mg, Ti و مقایسه پاسخ آن با دزیمتر (LiF: Mg, Ti (TLD-100 ساخت شرکت هارشا در برابر پرتوهای گاما
Thermoluminescence dosimeters are small tablets used with 3. 2*3. 2*0. 9mm^3 for measurement of received dose from radioactive various beams. The most common dosimeter is TLD-100 which is lithium-fluoride family and this dosimeter contain magnesium impurities. In this study, first lithium-flouride powder was mixed with titanium and controlled atmosphere samples were heated. After measuring ...
متن کاملevaluation of the entrance skin dose in animals undergoing diagnostic radiology using lif, mg, ti thermoluminescence dosimetry (tld-100)
introduction according to the international commission on radiological protection publication numbers 60 and 21, the environmental control standards should ensure human and other species protection to a desirable degree. since application of radiographic procedures in veterinary medicine has increased significantly, in this study, we aimed to evaluate the entrance skin dose to the animals (e.g....
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
journal of biomedical physics and engineeringجلد ۲۰۱۲، شماره ۱۲، صفحات ۰-۰
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023