application of ion mobility spectrometry for determination of morphine in human urine

نویسندگان

ali sheibani

m. reza shishehbore

shohrah vafaie-shahi

چکیده

in this study, a rapid, simple and sensitive ion mobility spectrometry (ims) method with corona discharge as ionization source was described for the morphine determination in human urine. morphine was extracted and purified from urine samples using solid phase extraction procedure with c18 column. it can offer the clean extracts which no extra peaks were observed in ims. under operating experimental conditions (temperature; injection: 220 and oven: 180 oc, flow rate; carrier: 300 and drift: 600 ml min-1,voltage;corona: 2200 and drift: 6700 v), developed method showed good linearity in the ranges of 0.44 to 6.91 and 6.91 to 22 ng ml-1 with correlation coefficients (r2) of 0.9979 and 0.9966, respectively. the limit of detection was 0.1 ng ml-1, and precision as relative standard deviation was 11%. the capability of the proposed method was evaluated by the analysis of human urine as a real sample that satisfactory results were obtained.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Application of Ion Mobility Spectrometry for Determination of Morphine in Human Urine

In this study, a rapid, simple and sensitive ion mobility spectrometry (IMS) method with corona discharge as ionization source was described for the morphine determination in human urine. Morphine was extracted and purified from urine samples using solid phase extraction procedure with C18 column. It can offer the clean extracts which no extra peaks were observed in IMS. Under operating experim...

متن کامل

application and construction of carbon paste modified electrodes developed for determination of metal ions in some real samples

ساخت الکترودهاِی اصلاح شده ِیکِی از چالشهاِی همِیشگِی در دانش شیمِی بوِیژه شیمِی تجزیه مِی باشد ،که با در نظر گرفتن سادگِی ساخت، کاربردی بودن و ارزان بودن روش مِی توان به باارزش بودن چنِین سنسورهاِی پِی برد.آنچه که در ادامه آورده شده به ساخت و کاربرد الکترودهاِی اصلاح شده با استفاده از نانو ذرات در اندازه گِیرِی ولتامترِی آهن وکادمِیم اشاره دارد. کار اول اختصاص دارد به ساخت الکترود خمِیر کربن اصلاح شده با لِیگاند داِ...

15 صفحه اول

Application of ion mobility spectrometry for the detection of human urine

The aim of the present study was to evaluate the suitability of ion mobility spectrometry (IMS) for the detection of human urine as an indication of human presence during urban search and rescue operations in collapsed buildings. To this end, IMS with a radioactive ionization source and a multicapillary column was used to detect volatile organic compounds (VOCs) emitted from human urine. A stud...

متن کامل

an application of equilibrium model for crude oil tanker ships insurance futures in iran

با توجه به تحریم های بین المملی علیه صنعت بیمه ایران امکان استفاده از بازارهای بین المملی بیمه ای برای نفتکش های ایرانی وجود ندارد. از طرفی از آنجایی که یکی از نوآوری های اخیر استفاده از بازارهای مالی به منظور ریسک های فاجعه آمیز می باشد. از اینرو در این پایان نامه سعی شده است با استفاده از این نوآوری ها با طراحی اوراق اختیارات راهی نو جهت بیمه گردن نفت کش های ایرانی ارائه نمود. از آنجایی که بر...

DETECTION AND MEASUREMENT OF ACETONE IN THE BREATH OF DIABETICS BY ION MOBILITY SPECTROMETRY METHOD

Background: The ion mobility spectrometry (IMS) is an analytical technique that is widely used due to its high sensitivity and speed for the detection of ionized molecules in gas phase and under atmospheric pressure. Breath analysis is a new method for obtaining information about person's clinical conditions that is considered by researchers. Human exhaled air contains a variety of components s...

متن کامل

Determination of Morphine and Codeine in Human Urine by Gas Chromatography-Mass Spectrometry

A sensitive and selective gas chromatography-mass spectrometry (GC-MS) method was developed and validated for the determination of morphine and codeine in human urine. The GC-MS conditions were developed. The analysis was carried out on a HP-1MS column (30 m × 0.25 mm, 0.25  μ m) with temperature programming, and Helium was used as the carrier gas with a flow rate of 1.0 mL/min. Selected ion mo...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید


عنوان ژورنال:
caspian journal of chemistry

ناشر: university of mazandaran

ISSN 1735-0611

دوره 2

شماره 1 2013

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023