measurement of intraocular pressure using schiotz vs tono-pen tonometers in patients with primary congenital glaucoma
نویسندگان
چکیده
purpose: to compare the tono-pen and schiotz tonometers in measurement of intraocular pressure (iop) in patients with primary congenital glaucoma. methods: this study was conducted on 38 eyes of 20 patients with primary congenital glaucoma. all measurements were performed under general anesthesia. one minute after receiving halothane 3% iop was determined by tono-pen and then by schiotz tonometers. data were analyzed by regression analysis and bland-altman bias plot. results: mean age of patients was 19.4±14.5 months and 15 (75%) of them were male. glaucoma was bilateral in 18 (90%) and unilateral in 2 (10%) subjects. mean iop measured by tono-pen and schiotz tonometers was 19.39±8.51 and 18.84±8.9 mmhg, respectively. these tonometers had good concordance in determination of iop in primary congenital glaucoma according to the results of regression analysis and bland-altman bias plot. conclusion: considering the sample size of this study, the tono-pen and schiotz tonometers are highly concordant regarding intraocular pressure measurement in children with primary congenital glaucoma.
منابع مشابه
Noninvasive measurement of rat intraocular pressure with the Tono-Pen.
PURPOSE The purpose of this study was to evaluate the Tono-Pen 2 tonometer for measuring intraocular pressure (IOP) in the living rat eye. METHODS One eye from each of 20 adult, anesthetized brown Norway rats (group 1) was cannulated and simultaneously connected to a syringe and a pressure transducer with a chart recorder. We increased IOP from 15 to 45 mmHg in 5-mmHg increments and obtained ...
متن کاملMeasurements of Intraocular Pressure with Tono-Pen and Goldmann Applanation Tonometers in Subjects with Systemic Illness
Objective: The purpose of this preliminary study was to compare differences in the measurement of intraocular pressure (IOP) between TonoPen tonometry and Goldmann applanation tonometry (GAT) in healthy individuals and those with coexisting illnesses. Patients and Methods: This prospective study included 42 individuals who were further categorized into two groups. Group 1 consisted of 20 health...
متن کاملIntraocular Pressure Measurement by Three Different Tonometers in Primary Congenital Glaucoma
PURPOSE To determine the agreement between intraocular pressure (IOP) measurements using an automated non-contact tonometer (NCT), Goldmann applanation tonometer (GAT), and the ocular response analyzer (ORA) in subjects with primary congenital glaucoma (PCG). METHODS Twenty-nine eyes of 17 PCG patients underwent IOP measurements using NCT, GAT and ORA. Variables obtained by the ORA were corne...
متن کاملIntraocular pressure measurement in cynomolgus monkeys. Tono-Pen versus manometry.
PURPOSE In living cynomolgus monkey eyes, the authors compared manometrically set and measured intraocular pressure (IOP) with simultaneous IOP readings obtained with the Tono-Pen (TP), a handheld applanation tonometer based on the Mackay-Marg principle. METHODS In three pentobarbital-anesthetized cynomolgus monkeys, IOP was set and measured manometrically after anterior chamber cannulation t...
متن کاملIntraocular Pressure Measurements by Three Different Tonometers in Children with Aphakic Glaucoma and a Thick Cornea
Background: To evaluate the agreement in intraocular pressure (IOP) measurements by Ocular Response Analyzer (ORA) and Tono-Pen XL (TXL) with the Goldmann Applanation Tonometer (GAT) and to examine corneal biomechanical properties in aphakic glaucoma patients with a central corneal thickness (CCT) >600 µ. Methods: Thirty-six eyes of aphakic glaucoma patients (group 1) and 40 eyes of normal chil...
متن کاملRepeatability of intraocular pressure measurements with Icare PRO rebound, Tono-Pen AVIA, and Goldmann tonometers in sitting and reclining positions
BACKGROUND Icare PRO (ICP) is a new Rebound tonometer that is able to measure intraocular pressure (IOP) in both sitting and reclining positions. In this study, the gold standard Goldmann tonometer (GAT) was compared to ICP and Tono-Pen AVIA (TPA). Hypothesis was that repeatability of GAT is superior to ICP and TPA. METHODS 36 eyes of 36 healthy caucasian individuals, 13 male and 26 females, ...