central corneal thickness measurements in normal and keratoconic eyes using three corneal pachymeters
نویسندگان
چکیده
purpose: to assess the agreement of ultrasonic pachymetry, the galilei dual scheimpflug analyzer and orbscan ii in measuring central corneal thickness (cct) in normal and keratoconic eyes. methods: in this prospective comparative study, cct was measured in 88 refractive surgery candidates and 76 (39 right) eyes of 45 keratoconic patients using ultrasonic pachymetry, galilei and orbscanii. readings by the three instruments were compared using one-way analysis of normal variance. agreement between the three devices was assessed using pearson (r) and intraclass correlation (icc) coefficients, mountain plots, and bland-altman plots. results: in the normal group, mean cct was 551.0±39.4, 566.9±33.5, and 565.5±40.9µm measured by ultrasonic pachymetry, galilei and orbscan ii, respectively (p<0.001). corresponding figures in the keratoconus group were 488.4±61.5, 504.1±42.0, and 473.8±58.5µm, respectively (p<0.001). cct measurements by the three instruments were significantly correlated in both groups. the association was strongest in the normal group and between the galilei and orbscanii (r=0.91, icc=0.90), where 95% limits of agreement (loa) were narrowest (-35 to 32 µm). the correlation was weakest in the keratoconus group and between ultrasonic pachymetry and galilei (r=0.67, icc=0.61). conclusion: to measure cct, galilei and orbscanǁ can be used interchangeably in normal eyes but not in keratoconic eyes. agreement between the two optical devices and ultrasonic pachymetry is low and therefore they cannot replace ultrasonic pachymetry in measuring cct in normal or keratoconic corneas.
منابع مشابه
Central and Peripheral Corneal Thickness Measurement in Normal and Keratoconic Eyes Using Three Corneal Pachymeters
PURPOSE To assess the agreement among ultrasonic pachymetry, the Galilei dual Scheimpflug analyzer, and Orbscan II for central and peripheral (Galilei vs. Orbscan) corneal thickness (CCT and PCT) measurement in normal and keratoconic eyes. METHODS In this prospective study, CCT and PCT were measured in 88 eyes of 88 refractive surgery candidates and 128 eyes of 69 keratoconic patients with ul...
متن کاملCorneal Biomechanical Assessment Using Corneal Visualization Scheimpflug Technology in Keratoconic and Normal Eyes
Purpose. To compare the corneal biomechanical properties of keratoconic patients and age-matched controls using corneal visualization Scheimpflug technology (Corvis ST). Methods. Sixty keratoconic eyes from 47 keratoconus patients and 60 normal eyes from 60 controls were enrolled in this prospective study. Tomography and biomechanical parameters of all eyes were obtained with the Pentacam and C...
متن کاملCorneal biomechanical properties in normal, regular astigmatic, and keratoconic eyes
Corneal hysteresis (CH) is a parameter which measures the viscoelastic behaviour of the cornea, indicating its biomechanical integrity1. Some clinical conditions such as keratoconus, Fuchs corneal dystrophy, high myopia, glaucoma and corneal refractive surgery, may induce changes in corneal biomechanical properties, leading to a decrease in CH2-6. Astigmatism, a refractive error related to corn...
متن کاملIntraocular Pressure Measurements Referring to the Corneal Thickness in Keratoconic Eyes After Corneal Crosslinking with Riboflavin and Ultraviolet A
AIM To determine the possible relation between intraocular pressure (IOP), central corneal thickness (CCT) and corneal resistance (CR) in kerotoconic eyes before, 3,6 and 12 months after collagen crosslinking procedure (CXL) with aim to find out does the thicker cornea means already more resistance cornea followed with higher IOP. METHODS Thirty eyes (30 patients) with central keratoconus (KC...
متن کاملAssessment of Corneal Epithelial Thickness in Asymmetric Keratoconic Eyes and Normal Eyes Using Fourier Domain Optical Coherence Tomography
Purpose. To compare the characteristics of asymmetric keratoconic eyes and normal eyes by Fourier domain optical coherence tomography (OCT) corneal mapping. Methods. Retrospective corneal and epithelial thickness OCT data for 74 patients were compared in three groups of eyes: keratoconic (n = 22) and normal fellow eyes (n = 22) in patients with asymmetric keratoconus and normal eyes (n = 104) i...
متن کاملCentral corneal thickness measurement with different devices in keratoconic patients
We read with great interest the article by Aghazadeh Amiri et al. concerning the corneal thickness (CT) measurements with Scheimpflug and slit scanning imaging techniques in keratoconus (KC). We would like to congratulate with the authors for their very interesting paper because it is well known that the measurement of CT is involved in the reliability of intraocular pressure measurements, both...