medium-term stability of the photon beam energy of an elekta compacttm linear accelerator based on daily measurements of beam quality factor

نویسندگان

mohammad amin mosleh-shirazi physics unit, department of radiotherapy and oncology, namazi hospital, shiraz university of medical sciences, shiraz, iran 2- ionizing and non-ionizing radiation protection research center, school of paramedical sciences, shiraz university of medical sciences, shiraz, iran

saeed rahimi department of medical physics and engineering, school of medicine, shiraz university of medical sciences, shiraz, iran student research committee, shiraz university of medical sciences, shiraz, iran

sareh karbasi physics unit, department of radiotherapy and oncology, namazi hospital, shiraz university of medical sciences, shiraz, iran

چکیده

introduction in this study, we aimed to assess the medium-term energy stability of a 6mv elekta compacttm linear accelerator. to the best of our knowledge, this is the first published article to evaluate this linear accelerator in terms of energy stability. as well as investigating the stability of the linear accelerator energy over a period of several weeks, the results will be useful for estimation of the required tolerance values for the beam quality factor (bqf) of the ptw quickcheck weblinetm (qcw) daily checking device. materials and methods over a 13 week period of routine clinical service, 52 daily readings of bqf were taken and then analyzed for a 10×10 cm2 field. results no decreasing or increasing trend in bqf was observed over the study period. the mean bqf value was estimated at 5.4483 with a standard deviation (sd) of 0.0459 (0.8%). the mean value was only 0.1% different from the baseline value. conclusion the results of this medium-term stability study of the elekta compact linear accelerator energy showed that 96.2% of the observed bqf values were within ±1.3% of the baseline value. this can be considered to be within the recommended tolerance for linear accelerator photon beam energy. if an approach of applying ±3 sd is taken, the tolerance level for bqf may be suggested to be set at ±2.5%. however, further research is required to establish a relationship between bqf value and the actual changes in beam energy and penetrative quality.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Medium-Term Stability of the Photon Beam Energy of An Elekta CompactTM Linear Accelerator Based on Daily Measurements of Beam Quality Factor

Introduction In this study, we aimed to assess the medium-term energy stability of a 6MV Elekta CompactTM linear accelerator. To the best of our knowledge, this is the first published article to evaluate this linear accelerator in terms of energy stability. As well as investigating the stability of the linear accelerator energy over a period of several weeks, the results will be useful for esti...

متن کامل

application of upfc based on svpwm for power quality improvement

در سالهای اخیر،اختلالات کیفیت توان مهمترین موضوع می باشد که محققان زیادی را برای پیدا کردن راه حلی برای حل آن علاقه مند ساخته است.امروزه کیفیت توان در سیستم قدرت برای مراکز صنعتی،تجاری وکاربردهای بیمارستانی مسئله مهمی می باشد.مشکل ولتاژمثل شرایط افت ولتاژواضافه جریان ناشی از اتصال کوتاه مدار یا وقوع خطا در سیستم بیشتر مورد توجه می باشد. برای مطالعه افت ولتاژ واضافه جریان،محققان زیادی کار کرده ...

15 صفحه اول

A Monte Carlo Simulation of Photon Beam Generated by a Linear Accelerator

ntroduction:  Monte  Carlo  simulation  is  the  most  accurate  method  of  simulating  radiation  transport  and  predicting doses at different points of interest in radiotherapy. A great advantage of the Monte Carlo method  compared  to  the  deterministic  methods  is  the  ability  to  deal  accurately  with  any  complex  geometry.  Its  disadvantage is the extremely long computing time r...

متن کامل

the effect of genre-based teaching on reading comprehension of literary texts

تحقیق حاضر به بررسی کاربرد روش ژانر-محور را در محیط آموزش زبان عمومی می پردازد.روش ژانر-محور به زبان آموزان کمک میکند که در زمینه خوانش پیشرفت کنند. بعضی از محققین معتقد اند که روش تدریس ژانر-محور به تدریج به زبان آموزان کمک می کند تا در درک ژانر های مختلف مهارت یابند (هایلند 2004).همچنین امروزه توجه روز افزونی به اهمیت استفاده از ادبیات در برنامه آموزشی زبان انگلیسی (esl/efl ) شده است. زمانی ک...

15 صفحه اول

the role of task-based techniques on the acquisition of english language structures by the intermediate efl students

this study examines the effetivenss of task-based activities in helping students learn english language structures for a better communication. initially, a michigan test was administered to the two groups of 52 students majoring in english at the allameh ghotb -e- ravandi university to ensure their homogeneity. the students scores on the grammar part of this test were also regarded as their pre...

15 صفحه اول

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید


عنوان ژورنال:
iranian journal of medical physics

جلد ۱۲، شماره ۴، صفحات ۲۳۰-۲۳۴

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023