a simple and sensitive hplc method for analysis of imipramine in human plasma with uv detection and liquid-liquid extraction: application in bioequivalence studies
نویسندگان
چکیده
high-performance liquid chromatography (hplc) methods employing ultraviolet (uv) detector are not sufficiently sensitive to measure the low plasma concentrations following single oral dose of imipramine. therefore, in the present study a simple, rapid and yet sensitive hplc method with uv detection was developed and validated for quantitation of imipramine in human plasma samples. an efficient liquid-liquid extraction (lle) of imipramine from plasma with the mixture of hexane/isoamyl alcohol (98:2) and back extraction of the drug in acidic medium concomitant with evaporation of organic phase allowed the use of uv detector to conveniently measure plasma levels of this compound as low level as 3 ng/ml. separation was achieved on a µ-bondapak c 18 hplc column using sodium hydrogen phosphate solution (0.01 m)/acetonitrile (60/40 v/v) at ph 3.5 ± 0.1 at 1.5 ml/min. trimipramine was used as the internal standard for analysis of plasma samples. the retention times for imipramine and trimipramine were 4.3 and 5.2 min, respectively. calibration curve was linear in the range of 3-40 ng/ml using human plasma with the average extraction recovery of 85 ± 5%. imipramine was found to be stable in plasma samples with no evidence of degradation during three freeze-thaw cycles and three months storage at -70 ºc. the current validated method was finally applied in bioequivalence studies of two different imipramine products according to a standard two-way crossover design with a two weeks washout period.
منابع مشابه
A simple and sensitive HPLC method for analysis of imipramine in human plasma with UV detection and liquid-liquid extraction: Application in bioequivalence studies
High-performance liquid chromatography (HPLC) methods employing ultraviolet (UV) detector are not sufficiently sensitive to measure the low plasma concentrations following single oral dose of imipramine. Therefore, in the present study a simple, rapid and yet sensitive HPLC method with UV detection was developed and validated for quantitation of imipramine in human plasma samples. An efficient ...
متن کاملSimple and Sensitive High-Performance Liquid Chromatography (HPLC) Method with UV Detection for Mycophenolic Acid Assay in Human Plasma. Application to a Bioequivalence Study.
PURPOSE A simple and available reversed-phase high performance liquid chromatography (HPLC) method with UV detection has been developed and validated for mycophenolic acid (MPA) assay in human plasma. METHODS MPA was extracted from plasma with protein precipitation method by acetonitrile: percholeric acid: methanol (75:5:20 v/v/v). The drug separation was achieved using a C8 analytical column...
متن کاملRapid High Performance Liquid Chromatographic Method for Determination of Clarithromycin in Human Plasma Using Amperometric Detection: Application in Pharmacokinetic and Bioequivalence Studies
A rapid, sensitive and reproducible HPLC method using amperometric detector was developed and validated for the analysis of clarithromycin in human plasma. The separation was achieved on a monolithic silica column (MZ- C8 125×4.0 mm) using acetonitrile- methanol-potassium dihydrogen phosphate buffer (40:6:54,v/v), with pH of 7.5, as the mobile phase at a flow rate of 1.5 mL/min. The assay enabl...
متن کاملRapid High Performance Liquid Chromatographic Method for Determination of Clarithromycin in Human Plasma Using Amperometric Detection: Application in Pharmacokinetic and Bioequivalence Studies
A rapid, sensitive and reproducible HPLC method using amperometric detector was developed and validated for the analysis of clarithromycin in human plasma. The separation was achieved on a monolithic silica column (MZ- C8 125×4.0 mm) using acetonitrile- methanol-potassium dihydrogen phosphate buffer (40:6:54,v/v), with pH of 7.5, as the mobile phase at a flow rate of 1.5 mL/min. The assay enabl...
متن کاملA quick and Sensitive Liquid Chromatography–tandem Mass Spectrometry (LC-MS) Method for the Determination of Enalapril and Enalaprilat in Human Plasma: Application to a Bioequivalence Study
A rapid and sensitive liquid chromatography–tandem mass spectrometry (LC-MS) method for the estimation of enalapril and enalaprilat in human plasma. Detection of analytes was achieved by tandem mass spectrometry with electrospray ionization (ESI) interface in positive ion mode was operated under the multiple-reaction monitoring mode. Sample pretreatment involved in a one-step protein precipitat...
متن کاملA simple and sensitive HPLC-UV method for quantitation of lovastatin in human plasma: application to a bioequivalence study.
An available, simple, sensitive, and rapid method has been developed for determination of the 3-hydroxy-3-methylglutaryl (HMG)-CoA reductase inhibitor, lovastatin in human plasma. The analytical procedure involves a one-step liquid-liquid extraction method using atorvastatin as internal standard. Chromatographic separation was carried out on a reversed phase C(18) column using a mixture of 0.05...