assessment of simulated patient motion and its effect on myocardial perfusion spect using two reconstruction methods (filtered backprojection;fbp and iterative method) [persian]
نویسندگان
چکیده
introduction: motion of the patient during myocardial perfusion spect could potentially results in false perfusion defects. the effect of different reconstruction methods on these artifacts is not studied. clarification of the relation between the extent, severity and duration of motion with the resultant artifacts may be helpful in designing special soft wares for motion correction. this study is a preliminary evaluation of motion artifacts in different reconstruction methods. materials and methods: normal myocardial perfusion spect from a patient with low risk of cad and no motion during imaging according to the review of cinematic images by 3 nuclear medicine specialists is selected. the scan was acquired one hour after iv injection of 740mbq of tc-99m-mibi through 180 degrees from rao 45 to lpo 45, in 32 frames and 30 seconds per frame. the original images were moved artificially in frames 7, 16 and 24 (as initial, mid and late portion of imaging) from 1 to 3 pixels and for 30-90 seconds using vision software. also this artificial motion was applied in x and y axis and in negative and positive directions. one hundred forty four images were produced and all were reconstructed using filtered back projection (fbp) and iterative (osem) reconstructions and interpreted visually and semi-quantitatively (using 17 segment model with 5 point scoring). results: applying one, two and three pixels motion converted normal scan to abnormal image in 2.8%, 25% and 39.6% respectively using fbp and in 18.8% , 43.2% and 68.8% respectively using osem reconstruction.(p
منابع مشابه
Assessment of simulated patient motion and its effect on myocardial perfusion SPECT using two reconstruction methods (Filtered Backprojection;FBP and Iterative method) [Persian]
Introduction: Motion of the patient during myocardial perfusion SPECT could potentially results in false perfusion defects. The effect of different reconstruction methods on these artifacts is not studied. Clarification of the relation between the extent, severity and duration of motion with the resultant artifacts may be helpful in designing special soft wares for motion correction. This...
متن کاملImage reconstruction using filtered backprojection and iterative method: effect on motion artifacts in myocardial perfusion SPECT.
UNLABELLED Patient motion during myocardial perfusion SPECT is a common source of errors. The extent and severity of motion artifacts have been described for filtered backprojection (FBP) reconstruction. In recent years, iterative reconstruction has been used increasingly in reconstruction of myocardial perfusion SPECT images and has been shown to be more accurate than FBP even in cases of inco...
متن کاملComparative study of myocardial perfusion SPECT and coronary angiography [Persian]
Abstract is NOT available
متن کاملtransference of imagery: a comparative formalistic study of shakespeares hamlet and its two persian translations
هدف از این تحقیق بررسی انتقال صور خیال هملت در دو ترجمه ی فارسی آن از نظر فرمالیستی بود. برای بدست آوردن داده-های مورد نیاز، 130 نمونه استعاره، مجاز، ایهام، کنایه و پارادوکس در متن اصلی مشخص شده و سپس بر اساس مدل نیومارک (1998) برای ترجمه ی استعاره یا بطور کلی زبان مجاز با معادل های فارسی شان مقایسه گردیدند. این تحقیق بر آن بود تا روش های استفاده شده برای ترجمه هر کدام از انواع زبان مجاز ذکر شد...
15 صفحه اولcomparative dna interaction studies of antiviral drug, zidovudine and its complex using different instrumental methods
هدف از این مطالعه بررسی امکان استفاده از داروهای شناخته شده در درمان سایر بیماریها به عنوان داروهای ضد سرطان است. همچنین با استفاده از این داروها در ساختمان کمپلکس فلز می توان شاخص های دارویی بدست آمده را بررسی نمود. داروی ضد ویروس ایدز(hiv)به نام زیدوودین(azt)انتخاب و.کمپلکس.محلول.در.آب[pt(azt)2]cl2سنتزو به روشهای مختلف فیزیکی و شیمیایی شناسایی گردید. بر هم کنش مقایسه ای این دارو و کمپلکس پلا...
15 صفحه اولMyocardial perfusion SPECT: Perfusion quantification
Different software tools for quantification of myocardial perfusion SPECT (MPS) studies are routinely used. Several perfusion parameters can be computed automatically. Interpretation of the MPS should start with visual inspection of the rotating planar images, visual analysis of reconstructed SPECT slices and then quantitative analysis to confirm the visual impression. Quantification should be...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
iranian journal of nuclear medicineناشر: tehran university of medical sciences
ISSN 1681-2824
دوره 13
شماره 1 2004
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023