comparison of cyclic fatigue resistance of five nickel titanium rotary file systems with different manufacturing techniques
نویسندگان
چکیده
objectives: the purpose of this study was to evaluate the resistance to fatigue failure of five different nickel-titanium rotary files in three different curved trajectories. materials and methods: a total of 150 neoniti a1, race 25.06, mtwo 25.06, twisted file 25.06 and protaper next x2 files with the tip size of 25 were tested (n=30 for each group). three groove types simulating curved canals were used differing in radius, arc length and position of the arc, each measuring 1.5 mm in width, 20 mm in total length and 2.5 in depth. resistance to cyclic fatigue was determined by counting the numbers of cycles to failure. furthermore, the fragment length of the fractured tips and angle and radius of curvature formed by each file in each trajectory were evaluated. the data were analyzed by t-test, one way anova and tukey’s hsd test. results: neoniti showed the highest and race showed the lowest number of cycles to fracture (ncf) values (p<0.05), indicating the highest and lowest fatigue resistance, respectively. the highest and lowest curvature angles were seen in race and neoniti, respectively. regarding the radius of curvature, twisted file had the lowest and neoniti had the highest values. the mean ncf of all rotary files was lower in the more coronally curved trajectory. conclusion: the fatigue resistance of the evaluated rotary files was lower in more coronally located curvatures. neoniti exhibited the highest and race exhibited the lowest fatigue resistance compared to other evaluated files.
منابع مشابه
Cyclic fatigue resistance and three-dimensional analysis of instruments from two nickel-titanium rotary systems.
AIM To determine how instrument design affects the fatigue life of two nickel-titanium (Ni-Ti) rotary systems (Mtwo and ProTaper) under cyclic fatigue stress in simulated root canals. METHODOLOGY Cyclic fatigue testing of instruments was performed in stainless steel artificial canals with radii of curvature of 2 or 5 mm and an angle of curvature of 60 degrees . A total of 260 instruments were...
متن کاملEffect of clinical use on the cyclic fatigue resistance of ProTaper nickel-titanium rotary instruments.
The resistance of ProTaper (Dentsply Maillefer, Ballaigues, Switzerland) nickel-titanium rotary instruments to cyclic fatigue was examined after their initial use in straight or curved canals in vivo. These instruments were rotated freely inside a steel phantom until separation. The number of rotations before failure and the lengths of the separated fragments were compared with data derived fro...
متن کاملCyclic fatigue resistance tests of Nickel-Titanium rotary files using simulated canal and weight loading conditions
OBJECTIVES This study compared the cyclic fatigue resistance of nickel-titanium (NiTi) files obtained in a conventional test using a simulated canal with a newly developed method that allows the application of constant fatigue load conditions. MATERIALS AND METHODS ProFile and K3 files of #25/.06, #30/.06, and #40/.04 were selected. Two types of testing devices were built to test their fatigu...
متن کاملCyclic Fatigue of Different Nickel-Titanium Rotary Instruments: A Comparative Study
Since the introduction of nickel-titanium alloy to endodontics, there have been many changes in instrument design, but no significant improvements in the raw material properties, or enhancements in the manufacturing process. Recently, a new method to produce nickel-titanium rotary (NTR) instruments has been developed, in an attempt to obtain instruments that are more flexible and resistant to f...
متن کاملCyclic Flexural Fatigue Resistance of the Revo-s Rotary Nickel-titanium Endodontic Files
Resistance to cyclic flexural fatigue failure is an important feature of rotary nickel-titanium (NiTi) instruments. The aim of this study was to evaluate the cyclic flexural fatigue failure resistance of RevoS Shaper Universal rotary NiTi files and to compare it to two other rotary NiTi files with similar size and taper, the Hero Shaper and the ProFile. Fifteen each of Revo-S Shaper Universal, ...
متن کاملInfluence of continuous or reciprocating motion on cyclic fatigue resistance of 4 different nickel-titanium rotary instruments.
INTRODUCTION The aim of this study was to evaluate the resistance to flexural fatigue of Reciproc R25 (VDW, Munich, Germany), WaveOne Primary (VDW, Munich, Germany), Mtwo (Sweden & Martina, Padova, Italy), and Twisted File (TF; SybronEndo, Orange, CA) instruments used in continuous rotation or in 2 different reciprocating motions. METHODS A total of 180 nickel-titanium files from 4 brands mar...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
journal of dentistry, tehran university of medical sciencesجلد ۱۲، شماره ۹، صفحات ۶۳۶-۶۴۶
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023