computed tomography colonography phantom: construction, validation and literature review

نویسندگان

lukas lambert department of radiology, first faculty of medicine, charles university in prague, prague, czech republic; department of radiology, general university hospital in prague, u nemocnice 2, 128 08 prague 2, czech republic. tel: +420-224962232, fax: +420-224963048

alena lambertova department of radiology, first faculty of medicine, charles university in prague, prague, czech republic; department of psychiatry, first faculty of medicine, charles university in prague, prague, czech republic

jan danes department of radiology, first faculty of medicine, charles university in prague, prague, czech republic

gabriela grusova fourth department of internal medicine, first faculty of medicine, charles university in prague, prague, czech republic

چکیده

conclusion we constructed and validated a ct colonography phantom with correct density allowing performance of reproducible experiments. materials and methods we found 15 different phantom designs by literature search and compared their advantages and disadvantages based on their description and images. we tested various materials for density and mechanical properties and constructed a phantom from polypropylene pipes (30 mm and 50 mm in diameter, 52 cm in length). haustral folds were created by heat shaping and 39 intermediate sessile polyps with a target size of 6-10 mm and two flat lesions were created from silicone. nine polyps were attached to a fold. the model was placed in a 30-cm barrel filled with water to simulate attenuation of human body. attenuation of polyps was compared to intermediate polyps found in patients. results none of the earlier colonic phantoms found in the literature incorporated all the properties that would ensure both reproducibility and validity of the model (including a rigid wall, density of the wall and polyps similar to human colon, at least two levels of distension and durability). in the present phantom, the average size of sessile polyps was 8.6 ± 0.9 mm and their density was 53 ± 24 hu. we found no significant difference in polyp density between simulated polyps in the phantom and polyps in human subjects (p = 0.70). all polyps, with the exception of one flat lesion, were detected by computer aided detection. objectives to compare their pros and cons and describe the construction of a dedicated phantom that can be easily manufactured. background several dedicated computed tomography (ct) colonography phantoms have been described previously.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Computed Tomography Colonography Phantom: Construction, Validation and Literature Review

BACKGROUND Several dedicated computed tomography (CT) colonography phantoms have been described previously. OBJECTIVES To compare their pros and cons and describe the construction of a dedicated phantom that can be easily manufactured. MATERIALS AND METHODS We found 15 different phantom designs by literature search and compared their advantages and disadvantages based on their description a...

متن کامل

Design and Construction of an Affordable Phantom for Electron Density Measurement and Linearity Tests of Computed Tomography Systems

Introduction: The performance of computed tomography is routinely checked using phantoms, which are known as important diagnostic imaging tools. Depending upon the aim of the study, different phantoms are designed, while trying to satisfy certain levels of diversities in their application. Material and Methods: The present study describes the...

متن کامل

[Computed tomography colonography image assessment using colon phantom-effects of reconstructed slice thickness].

Computed tomography colonography (CTC) is a robust and reliable imaging test of the colon. Recent studies show good sensitivity for the identification of nonpolypoid (flat) lesions as well. The purpose of this study was to determine the accuracy and reproducibility of a volume-rendering (VR) and virtual gross pathology (VGP) technique for detecting a polypoid lesion phantom by varying slice thi...

متن کامل

Computed tomography colonography for colorectal cancer screening.

Colorectal cancer screening has been shown to help prevent cancer-related death. Concerns about suboptimal adherence to conventional colonoscopy has led to the search for alternative screening modalities. Computed tomography colono-graphy (CTC) is a highly sensitive and minimally invasive alternative modality. The American College of Radiology has established a standardized reporting system for...

متن کامل

Validation of computed tomography-based attenuation correction of deviation between theoretical and actual values for four computed tomography scanners

Objective: In this study, we aimed to validate the accuracy of computed tomography-based attenuation correction (CTAC) using the bilinear scaling method.Methods: The measured attenuation coefficient (μm) was compared to a theoretical attenuation coefficient (μt ) using four different CT scanners and an RMI 467 phantom. The effective energy of the CT beam X-rays was calculated, using the aluminu...

متن کامل

Computed tomography colonography – reasons for different and false results

Computed tomography colonography (CT colonography) is one of the latest radiological methods of colorectal diagnostic imaging. Many studies confirmed a high efficacy of CT colonography in diagnosing colorectal polyps and tumors. However, this imaging method is not devoid of false diagnoses. Our paper presented the main causes of false results, causes of heterogeneity of the results among centre...

متن کامل