the effect of inversion times on the minimum signal intensity of the contrast agent concentration using inversion recovery t1-weighted fast imaging sequence
نویسندگان
چکیده
background :inversion recovery (ir) pulse sequences can generate t1-weighted images with a different range of inversion time (ti) to suppress or null the signal intensity (si) for a specified tissue. in this study, we aimed to investigate the effect of ti values on the concentration of the contrast agent, which leads to a minimum signal intensity, using an inversion recovery t1-weighted 3-dimensional fast-gradient echo imaging sequence. methods : a phantom was designed to hold 25 vials which contained different (between 0 and 19.77mmol/l of (gd-dtpa)) concentrations of the contrast agent. we used the vials of different concentrations to measure si using ir sequences with different inversion times (ti, 100-3000ms). results : the results of this study revealed that the t1 recovery curve did not cross the x- axis for the lower ti. therefore, a minimum si can be observed from the concentration of the contrast agent versus si curves. the findings of this study also revealed that the concentration of the contrast agent, which leads to a minimum si, is dependent on the ti and the minimum si will be increased at higher ti concentrations. conclusion : in conclusion, when the ti parameter is used to suppress the si of the specified tissues in clinical studies (e.g., fat suppression or blood suppression in perfusion measurements), it should be chosen with great caution.
منابع مشابه
The effect of inversion times on the minimum signal intensity of the contrast agent concentration using inversion recovery t1-weighted fast imaging sequence
Background :Inversion recovery (IR) pulse sequences can generate T1-weighted images with a different range of inversion time (TI) to suppress or null the signal intensity (SI) for a specified tissue. In this study, we aimed to investigate the effect of TI values on the concentration of the contrast agent, which leads to a minimum signal intensity, using an inversion recovery T1-weighted 3-dim...
متن کاملThe effect of inversion times on the minimum signal intensity of the contrast agent concentration using inversion recovery t1-weighted fast imaging sequence
BACKGROUND Inversion recovery (IR) pulse sequences can generate T1-weighted images with a different range of inversion time (TI) to suppress or null the signal intensity (SI) for a specified tissue. In this study, we aimed to investigate the effect of TI values on the concentration of the contrast agent, which leads to a minimum signal intensity, using an inversion recovery T1-weighted 3-dimens...
متن کاملcomparison of maximum signal intensity of contrast agent on t1-weighted images using spin echo, fast spin echo and inversion recovery sequences
conclusion using standard imaging parameters, each mri sequence reaches its maximum si in a specific contrast concentration, which is highest in se and least in ir in a comparison between se, fse and ir sequences. results this study shows that maximum si will appear at different ranges in different sequences. the maximum si can be seen at concentrations of 5.95, 4.96 and 3.98mmol/l for se, fse ...
متن کاملEffect of Echo Time on the Maximum Relationship between Contrast Agent Concentration and Signal Intensity Using FLAIR Sequence
Introduction Contrast-enhanced fluid-attenuated inversion recovery (FLAIR) is one of the MRI sequences that can be used for detection and evaluation of pathological changes in the brain. In this work, we have studied the effect of different echo times (TE) on the maximum relationship between signal intensity and concentration of the contrast agent using the FLAIR sequence. Materials and Methods...
متن کاملthe effect of using model essays on the develpment of writing proficiency of iranina pre-intermediate efl learners
abstract the present study was conducted to investigate the effect of using model essays on the development of writing proficiency of iranian pre-intermediate efl learners. to fulfill the purpose of the study, 55 pre- intermediate learners of parsa language institute were chosen by means of administering proficiency test. based on the results of the pretest, two matched groups, one as the expe...
effect of flip angle on the correlation between signal intensity and different concentrations of iron oxide nanoparticles using t1-weighted turbo-flash inversion recovery sequence
results the maximum si was obtained at the highest applied flip angle (45°). the linear relationship between si and nanoparticle concentration was seen up to 112.21 and 98.83 μmol fe/l for the short (10°) and the long (45°) flip angles, respectively (r2 = 0.95). these values were reduced up to 48.54 and 42.73 μmol fe/l for these flip angles with r2 of 0.99. conclusions the maximum si will be in...