beam hardening artifacts: comparison between two cone beam computed tomography scanners
نویسندگان
چکیده
background and aims. at present, cone beam computed tomography (cbct) has become a substitute for computed tomography (ct) in dental procedures. the metallic materials used in dentistry can produce artifacts due to the beam hardening phenomenon. these artifacts decrease the quality of images. in the present study, the number of artifacts as a result of beam hardening in the images of dental implants was compared between two newtom vg and planmeca promax 3d max cbct machines. materials and methods. an implant drilling model was used in the present study. the implants (dentis) were placed in the canine, premolar and molar areas. scanning procedures were carried out by two cbct machines. the corresponding sections (coronal and axial) of the implants were evaluated by two radiologists. the number of artifacts in each image was determined using the scale provided. mann-whitney u test was used for two-by-two comparisons at a significance level of p<0.05. results. there were statistically significant differences in beam hardening artifacts in axial and coronal sections between the two x-ray machines (p<0.001), with a higher quality in the images produced by the newtom vg. conclusion. given the higher quality of the images produced by the newtom vg x-ray machine, it is recommended for imaging of patients with extensive restorations, multiple prostheses or previous implant treatments.
منابع مشابه
Beam Hardening Artifacts: Comparison between Two Cone Beam Computed Tomography Scanners
BACKGROUND AND AIMS At present, cone beam computed tomography (CBCT) has become a substitute for computed tomography (CT) in dental procedures. The metallic materials used in dentistry can produce artifacts due to the beam hard-ening phenomenon. These artifacts decrease the quality of images. In the present study, the number of artifacts as a result of beam hardening in the images of dental imp...
متن کاملbeam hardening artifacts by dental implants: comparison of cone‑beam and 64‑slice computed tomography scanners
background: cone beam computed tomography (cbct) is an alternative to a computed tomography (ct) scan, which is appropriate for a wide range of craniomaxillofacial indications. the long‑term use of metallic materials in dentistry means that artifacts caused by metallic restorations in the oral cavity should be taken into account when utilizing cbct and ct scanners. the aim of this study was to ...
متن کامل[Comparison of absorbed and effective dose from two dental cone beam computed tomography scanners].
OBJECTIVE This research compared the field of view with the tissue absorbed dose and effective doses using the two dental cone beam computed tomography (CBCT) scanners. MATERIALS AND METHODS Two CBCT devices, an Alphard VEGA and 3DX multi-image micro CT MCT-1, were used. Measurements were made using an Alderson RANDO phantom and thermoluminescence dosimeters (TLDs). The calculation of the eff...
متن کاملEffective dose range for dental cone beam computed tomography scanners.
OBJECTIVE To estimate the absorbed organ dose and effective dose for a wide range of cone beam computed tomography scanners, using different exposure protocols and geometries. MATERIALS AND METHODS Two Alderson Radiation Therapy anthropomorphic phantoms were loaded with LiF detectors (TLD-100 and TLD-100 H) which were evenly distributed throughout the head and neck, covering all radiosensitiv...
متن کاملComparison of Spiral Computed Tomography and Cone-Beam Computed Tomography
Imaging is a highly dependent technology in the clinical assessment of a patient. With respect to the dental field, imaging has travelled a long way from the conventional radiographs to modern techniques like the computed tomography (CT) and the cone-beam computed tomography (CBCT). This article describes the evolution made in the CT in the field of dental imaging and attempts to compare the CT...
متن کاملArtifacts of Titanium, Zirconium, and Binary Titanium-Zirconium Abutments in Compute Tomography, Cone Beam Computed Tomography, and Magnetic Resonance Imaging
Aim: The aim of this in vitro study was to evaluate imaging artifacts induced by Titanium, Zirconium, Titanium-Zirconium abutments in CT, MRI and CBCT imaging modalities.Methods: A 4×8mm titanium fixture was inserted in a dry human mandible. Titanium, Zirconium and Titanium–Zirconium abutments measuring 10.5 mm in height are located on the fixture, one by one. Each abutment was scan...