comparison of prophylactic effects of polyurethane cylindrical or tapered cuff and polyvinyl chloride cuff endotracheal tubes on ventilator-associated pneumonia.
نویسندگان
چکیده
because microaspiration of contaminated supraglottic secretions past the endotracheal tube cuff is considered to be central in the pathogenesis of pneumonia, improved design of tracheal tubes with new cuff material and shape have reduced the size and number of folds, which together with the addition of suction ports above the cuff to drain pooled subglottic secretions leads to reduced aspiration of oropharyngeal secretions. so we conducted a study to compare the prophylactic effects of polyurethane-cylindrical or tapered cuff and polyvinyl chloride cuff endotracheal tubes (ett) on ventilator-associated pneumonia. this randomized clinical trial was carried out in a 12 bed surgical intensive care unit. 96 patients expected to require mechanical ventilation more than 96 hours were randomly allocated to one of three following groups: polyvinyl chloride cuff (pcv) ett, polyurethane (pu) cylindrical sealguard ett and pu taperguard ett. cuff pressure monitored every three hours 3 days in all patients. mean cuff pressure didn't have significant difference between three groups during 72 hours. pneumonia was seen in 11 patients (34%) in group pvc, 8 (25%) in sealguard and 7 (21%) in taperguard group. changes in mean cuff pressure between sealguard and pvc tubes and also between taperguard and pvc tubes did not show any significant difference. there was no significant difference in overinflation between three groups. the use of ett with pu material results in reducing ventilator-associated pneumonia compared to ett with pvc cuff. in pu tubes taperguard has less incidence of ventilator-associated pneumonia compared to sealguard tubes.
منابع مشابه
Comparison of prophylactic effects of polyurethane cylindrical or tapered cuff and polyvinyl chloride cuff endotracheal tubes on ventilator-associated pneumonia.
Because microaspiration of contaminated supraglottic secretions past the endotracheal tube cuff is considered to be central in the pathogenesis of pneumonia, improved design of tracheal tubes with new cuff material and shape have reduced the size and number of folds, which together with the addition of suction ports above the cuff to drain pooled subglottic secretions leads to reduced aspiratio...
متن کاملComparison of a novel lycra endotracheal tube cuff to standard polyvinyl chloride cuff and polyurethane cuff for fluid leak prevention.
BACKGROUND A high-volume low-pressure endotracheal tube (ETT) cuff forms folds along its contact with the trachea, allowing mucus leakage into the lungs. We developed a thin-walled ETT cuff made of Lycra polyurethane. METHODS In vitro, we tested 6 of each of the new prototype Lycra cuff, the Mallinkrodt Hi-Lo ETT (polyvinyl chloride cuff), and the Kimberly-Clark Microcuff ETT (polyurethane cu...
متن کاملThe impact of hospital-wide use of a tapered-cuff endotracheal tube on the incidence of ventilator-associated pneumonia.
BACKGROUND Aspiration of colonized oropharyngeal secretions is a major factor in the pathogenesis of ventilator-associated pneumonia (VAP). A tapered-cuff endotracheal tube (ETT) has been demonstrated to reduce aspiration around the cuff. Whether these properties are efficacious in reducing VAP is not known. METHODS This 2-period, investigator-initiated observational study was designed to ass...
متن کاملDo endotracheal tubes with suction above the cuff decrease the rate of ventilator-associated pneumonia, and are they cost-effective?
1. Department of Pulmonology, Faculdade de Medicina, Universidade de São Paulo USP São Paulo (SP), Brazil. 2. Faculdade de Medicina, Universidade de São Paulo USP São Paulo (SP), Brazil. Endotracheal tube-associated pneumonia and ventilator-associated pneumonia (VAP) occur in approximately 10% to 25% of patients who require invasive mechanical ventilation for more than 24 hours; the incidence d...
متن کاملImpact of supra-cuff suction on ventilator-associated pneumonia prevention
Critically ill patients are intubated or tracheostomized because, in most cases, these individuals require invasive mechanical ventilation. The cannulae that are used include the cuff, which can act as a reservoir for oropharyngeal secretions, predisposing to ventilator-associated pneumonia. Studies have revealed that the suction of subglottic secretions through the dorsal suction lumen above t...
متن کاملA quantitative evaluation of fluid leakage around a polyvinyl chloride tapered endotracheal tube cuff using an in-vitro model
INTRODUCTION This study was designed to quantitatively evaluate the sealing effect of a polyvinyl chloride tapered endotracheal tube cuff and to compare the tapered cuff with cylindrical endotracheal tube cuffs using an in vitro viscous fluid model. METHODS Five types of 8.0 mm inner diameter endotracheal tubes (TaperGuard, PortexSacett, PortexSoftseal, Sheridan HVT, Sheridan CF) were fixed i...