new approach for calibration of pixelated scintillation detectors of intraoperative gamma cameras
نویسندگان
چکیده
introduction:surgeosight is a dedicated intra-operative hand-held gamma camera designed and manufactured by our team to be used for lymphoscintigraphy. although conventional gamma cameras are widely used, they lack the capability of the sentinel lymph node (sln) imaging in the operation room. like conventional gamma cameras it is necessary to calibrate surgeosight for linearity distortion, energy and flood correction. our main objective in this study is to develop and implement three techniques for linearity, energy and uniformity corrections. methods: the camera is made up of a pixelated cesium iodide (sodium-activated) (csi(na)) scintillation crystal with 1 × 1 mm2 element size along with a hamamatsu h8500 flat-panel multi-anode (64 channels) photomultiplier tube. all triple corrections required lookup tables (luts). to generate all three luts, one uncalibrated acquisition is necessary. linearity distortion correction was based on localization of physical pixels of crystal and thin plate spline interpolation to estimate the amount of distortion. energy correction lut was simply a photo-peak map of incoming events which was estimated using self-organizing map technique. results: the results showed improvements qualitatively and quantitatively. the extrinsic energy resolution was enhanced from 50% to about 20.6%. integral uniformity and the differential uniformity, after uniformity correction, in useful field-of-view (ufov) was measured 9.5% and 4.5%, respectively. conclusion: results substantiated that the correction techniques guarantee uniform and accurate output of the surgeosight which is desirable for intra-operative localization of the sentinel lymph nodes in breast cancer and also for imaging of other superficial tumors.
منابع مشابه
New approach for calibration of pixelated scintillation detectors of intraoperative gamma cameras
Introduction:SURGEOSIGHT is a dedicated intra-operative hand-held gamma camera designed and manufactured by our team to be used for lymphoscintigraphy. Although conventional gamma cameras are widely used, they lack the capability of the sentinel lymph node (SLN) imaging in the operation room. Like conventional gamma cameras it is necessary to calibrate SURGEOSIGHT for linearity...
متن کامل A novel approach to automatic position calibration for pixelated crystals in gamma imaging
Introduction: The position estimation in gamma detection system will have constant misplacements which can be corrected in the calibration procedure. In the pixelated crystal uniformly irradiation of detector will produce irregular shape due to position estimation errors. This image is called flood field image and is used to calibrate the position estimation. In this work we pr...
متن کاملthe use of appropriate madm model for ranking the vendors of mci equipments using fuzzy approach
abstract nowadays, the science of decision making has been paid to more attention due to the complexity of the problems of suppliers selection. as known, one of the efficient tools in economic and human resources development is the extension of communication networks in developing countries. so, the proper selection of suppliers of tc equipments is of concern very much. in this study, a ...
15 صفحه اولMaximum-likelihood scintillation detection for EM-CCD based gamma cameras.
Gamma cameras based on charge-coupled devices (CCDs) coupled to continuous scintillation crystals can combine a good detection efficiency with high spatial resolutions with the aid of advanced scintillation detection algorithms. A previously developed analytical multi-scale algorithm (MSA) models the depth-dependent light distribution but does not take statistics into account. Here we present a...
متن کاملA Unified Calibration Approach for Generic Cameras
The classic perspective projection is mostly used when calibrating a camera. Although this approach is fairly developed and often suitable, it is not necessarily adequate to model any camera system like fish-eyes or catadioptrics. The perspective projection is not applicable when field of views reach 180° and beyond. In this case an appropriate model for a particular non perspective camera has ...
متن کاملScintillation detectors for x-rays
Recent research in the field of phosphor and scintillator materials and related detectors is reviewed. After a historical introduction the fundamental issues are explained regarding the interaction of x-ray radiation with a solid state. Crucial parameters and characteristics important for the performance of these materials in applications, including the employed measurement methods, are describ...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
iranian journal of nuclear medicineجلد ۲۵، شماره ۱، صفحات ۳۴-۴۲
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023