determination of trace amounts of lead by adsorptive cathodic stripping voltammetry in the presence of xylenol orange
نویسندگان
چکیده
lead forms the complexes with xylenol orange in the basic solution. an adsorptivedifferential pulse stripping method for the determination of lead is proposed. the procedureinvolves an adsorptive accumulation of lead on a hanging mercury drop electrode (hmde) inthe presence of xylenol orange, followed by reduction of adsorbed lead by voltammetric scanusing differential pulse modulation. the optimum experimental conditions are: xylenol orangeconcentration of 0.20 μg ml-1, ph 10.5, accumulation potential of -900 mv versus ag/agcl,accumulation time of 200 s, scan rate of 10 mv s-1 and pulse height of 100 mv. the peak currentis proportional to the concentration of lead over range of 0.03-0.65 μg ml-1, and the detectionlimit is 0.01 μg ml-1. the relative standard deviations (n=3) for 0.10 μg ml-1 lead is 0.86%,respectively. the proposed method was applied to the determination of lead in hair and watersamples with satisfactory results.
منابع مشابه
Determination of trace amounts of lead by adsorptive cathodic stripping voltammetry in the presence of xylenol orange
Lead forms the complexes with xylenol orange in the basic solution. An adsorptivedifferential pulse stripping method for the determination of lead is proposed. The procedureinvolves an adsorptive accumulation of lead on a hanging mercury drop electrode (HMDE) inthe presence of xylenol orange, followed by reduction of adsorbed lead by voltammetric scanusing differential pulse modulation. The opt...
متن کاملDetermination of trace amounts of thallium by adsorptive cathodic stripping voltammetry with xylenol orange.
Trace amounts of thallium(I) can be determined using adsorptive cathodic stripping voltammetry in the presence of Xylenol Orange (XO). The reduction current of the thallium(I)-XO complex ion was measured by square-wave cathodic stripping voltammetry. The peak potential was at -0.44 V vs. Ag/AgCl. The effect of various parameters (pH, ligand concentration, accumulation potential and collection t...
متن کاملSimultaneous Determination of Trace Amounts of Lead and Zinc by Adsorptive Cathodic Stripping Voltammetry
A selective and sensitive method for simultaneous determination of lead and zinc by adsorptive differential pulse cathodic stripping voltammetry is presented. The method is based on adsorptive accumulation of the complexes of Pb (II) and Zn(II) ions with Dopamine onto hanging mercury drop electrode (HMDE), followed by reduction of adsorbed species by differential pulse cathodic stripping voltam...
متن کاملDetermination of Trace Amounts of Nickel by Differential Pulse Adsorptive Cathodic Stripping Voltammetry Using Calconcarboxylic Acid as a Chelating Agent
A method for the determination of nickel(II) by stripping voltammetry is described. The method is based on the adsorptive accumulation of nickel(II) calconcarboxylic acid complex onto a hanging mercury drop electrode (HMDE), followed by the reduction of the adsorbed complex using differential pulse voltammetry. The optimum operating conditions and parameters were found to be 0.05 M NH3/NH4Cl bu...
متن کاملSimultaneous determination of copper, bismuth and lead by adsorptive stripping voltammetry in the presence of thymolphthalexone.
A novel, sensitive and selective adsorptive stripping procedure for simultaneous determination of copper, bismuth and lead is presented. The method is based on the adsorptive accumulation of thymolphthalexone (TPN) complexes of these elements onto a hanging mercury drop electrode, followed by reduction of adsorbed species by voltammetric scan using differential pulse modulation. The influences ...
متن کاملDetermination of Aminophylline by Cathodic Stripping Voltammetry
Aminophylline is determined by cathodic stripping voltammetry (CSV) in BR buffer, pH 7.5 at a hanging mercury drop electrode. The detection limit was 3x10 M after 60 s accumulation at -0.60 V versus Ag/AgCl reference electrode. The linear range demonstrated up to 5x10 M using CSV. The interference caused by some purine derivatives, anions and some metal cations on the peak current was studied. ...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
journal of the iranian chemical researchجلد ۳، شماره ۴، صفحات ۲۷۱-۲۷۸
کلمات کلیدی