utilization of electrodeposition on a graphite probe modified with palladium in determination of lead by graphite furnace atomic absorption spectrometry in water and environmental samples

نویسندگان

ahmad rouhollahi

department of chemistry, faculty of science, k.n. toosi university of technology, p.o. box 16315- 1618, tehran, iran reza moradkhani

department of chemistry, faculty of science, k.n. toosi university of technology, p.o. box 16315- 1618, tehran, iran hamid shirkhanloo

department of chemistry, faculty of science, k.n. toosi university of technology, p.o. box 16315- 1618, tehran, iran jahan b. ghasemi

department of chemistry, faculty of science, k.n. toosi university of technology, p.o. box 16315- 1618, tehran, iran

چکیده

in this work a rapid and selective procedure for separation and preconcentration of lead(pb2+) before determination by electrothermal atomic absorption spectrometry (etaas) wasdeveloped. the procedure is based on the electrodeposition on a graphite probe modified withpalladium. the lead was deposited from acetate buffer solution at ph 5.5. various parameters,such as ph of solutions, deposition potential, buffer concentration, stirrer speed, time ofdeposition and temperature program, were optimized. after optimization of the conditions,detection limit 17 ng l–1 by 3 σ, and enrichment factor 61 were achieved for 2 minelectrodeposition time and improved as deposition time was increased. linearity of calibrationwas kept between 0.05-0.50 μg l–1 with a correlation coefficient of 0.9979 and suitableprecision, r.s.d. % = 5.1 (n = 8). samples were digested completely in a closed microwavedigestion system using only perchloric acid, and interference owing to various cations was alsoinvestigated. the procedure was successfully applied to determine the presence of lead in rice,radish, okra, onion and water samples.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Utilization of electrodeposition on a graphite probe modified with palladium in determination of lead by graphite furnace atomic absorption spectrometry in water and environmental samples

In this work a rapid and selective procedure for separation and preconcentration of lead(Pb2+) before determination by electrothermal atomic absorption spectrometry (ETAAS) wasdeveloped. The procedure is based on the electrodeposition on a graphite probe modified withpalladium. The lead was deposited from acetate buffer solution at pH 5.5. Various parameters,such as pH of solutions, deposition ...

متن کامل

GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROMETRY WITH MAlIlRIX MODIFICATION FOR DETERMINATION OF LEAD IN DRINKING WATER

An amoniumvanadate plus sodium molybdate mixed modifier Graphite Furnace Atomic Absorption Spectrometric method has been developed for determination of traces of lead in drinking water. The concentration of lead in the sample is obtained by direct comparison to linear working graph prepared from aqueous standards in the modifier medium, The results indicate the better performance of the pro...

متن کامل

graphite furnace atomic absorption spectrometry with malilrix modification for determination of lead in drinking water

an amoniumvanadate plus sodium molybdate mixed modifier graphite furnace atomic absorption spectrometric method has been developed for determination of traces of lead in drinking water. the concentration of lead in the sample is obtained by direct comparison to linear working graph prepared from aqueous standards in the modifier medium, the results indicate the better performance of the propose...

متن کامل

Determination of lead in croatian wines by graphite furnace atomic absorption spectrometry.

A method has been developed for direct determination of lead in wine by graphite furnace atomic absorption spectrometry (GFAAS) with Zeeman-effect background correction. The thermal behaviour of Pb during pyrolysis and atomisation stages was investigated without matrix modifier and in the presence of Pd(NO(3))(2), Pd(NO(3))(2) + Mg(NO(3))(2) × 6H(2)O, and NH(4)H(2)PO(4) + Mg(NO(3))(2) × 6H(2)O ...

متن کامل

Automated on-line separation preconcentration system for palladium determination by graphite furnace atomic absorption spectrometry and its application to palladium determination in environmental and food samples.

A flow injection (FI) system was used to develop an efficient on-line sorbent extraction preconcentration system for palladium by graphite furnace atomic absorption spectrometry (GFAAS). The investigated metal was preconcentrated on a microcolumn packed with 1,5-bis(di-2-pyridyl)methylene thiocarbohydrazide immobilized on silica gel (DPTH-gel). The palladium is eluted with 40mul of HCl 4M and d...

متن کامل

DEVELOPMENT OF A METHOD FOR ALUMINIUM DETERMINATION IN SERUM AND DIALYSIS FLUID BY FLAMELESS ATOMIC ABSORPTION WITH GRAPHITE FURNACE

Aluminium determination was carried out in serum and dialysis fluid by a simple and reliable method of nameless atomic absorption with graphite furnace . . No preparatory procedures are required for water and dialysis fluid. In this method serum was mixed with 0.2% HNO, to allow complete combustion of the samples in order to improve analytical precision. The method has a sensitivity of 15 p...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید


عنوان ژورنال:
journal of the iranian chemical research

جلد ۴، شماره ۳، صفحات ۱۵۳-۱۶۴

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023