measurement of organ dose in abdomen-pelvis ct exam as a function of ma, kv and scanner type by monte carlo method .
نویسندگان
چکیده
background: ct is a diagnostic imaging modality giving higher patient dose in comparison with other radiological procedures, so the calculation of organ dose in ct exams is very important. while methods to calculate the effective dose have been established (icrp 26 and icrp 60), they depend heavily on the ability to estimate the dose to radiosensitive organs from the ct procedure. however, determining the radiation dose to these organs is problematic, direct measurement is not possible and comparing the dose as functions of scan protocol such as ma is very difficult. one of the most powerful tools for measuring the organ dose is monte carlo simulation. materials and methods: today the predominant method for assessment of organ absorbed dose is the application of conversion coefficients established by the use of monte carlo simulations. one of the most famous dose calculation software is ctdose, which we have used it for calculation of organ dose. in this work we measured the relationship between the ma, kv and scanner type with the equivalent organ dose and effective dose in mathematically standard phantom (hermaphrodite 170cm/70kg) in an abdomen-pelvis ct exam by monte carlo method. for this measurement we increased the ma in steps of 10 ma and plot curves for organ dose as a function of ma for different kv setting. results: as expected, with increasing ma, patient organ dose increased, but the simulation results showed that the slope of organ dose as a function of ma increased with kv increasing. by increasing kv from 120 to 140 the increase in slope of curves representing patient organ dose versus ma for different scanner types show almost similar behavior whereas the slope of the corresponding curves in scanners which equipped xenon detectors was almost 22% more than the slope of scanners equipped with scintillation detectors. conclusion: our research showed that regarding equivalent dose the system incorporating scintillation detector has a superior performance. incorporating such software in various ct scanners, marketed by different vendors, will offer the ability to get a print out of patient organ dose in any examination according to the imaging parameters used for imaging any part of the body. iran . j. radiat. res., 2004 1(4): 187-194
منابع مشابه
Measurement of organ dose in abdomen-pelvis CT exam as a function of mA, KV and scanner type by Monte Carlo method .
Background: CT is a diagnostic imaging modality giving higher patient dose in comparison with other radiological procedures, so the calculation of organ dose in CT exams is very important. While methods to calculate the effective dose have been established (ICRP 26 and ICRP 60), they depend heavily on the ability to estimate the dose to radiosensitive organs from the CT procedure. However, dete...
متن کاملAbsorbed dose of Fetus in SPECT-CT Imaging of Mother's Abdomen Regions using Monte Carlo method
SPECT-CT is an imaging method for diagnosing diseases and studying physiology of the body. In this type of imaging, due to the use of radiopharmaceutical for imaging (SPECT) and X-rays (CT), a relatively high dose of radiation is introduced into the patient. If the patient is pregnant, considering that the developing fetus is sensitive to any radiation in the uterus throughout the entire period...
متن کاملMeasurement of organ dose in chest CT examination using Monte Carlo simulation
Background: Exposure conditions in CT examinations are quite different from conventional Xray. In CT examination higher dose is given to patient in comparison with the dose in other diagnostic examinations. In order to calculate organ effective dose in chest CT, Monte Carlo simulation has been used in this study. Materials and Methods: The Impact survey data were used to determine the ...
متن کاملmeasurement of organ dose in chest ct examination using monte carlo simulation
background: exposure conditions in ct examinations are quite different from conventional xray. in ct examination higher dose is given to patient in comparison with the dose in other diagnostic examinations. in order to calculate organ effective dose in chest ct, monte carlo simulation has been used in this study. materials and methods: the impact survey data were used to determine the parameter...
متن کاملMonte-Carlo Calculation of Radon Absorbed Dose in Optical Fiber as a Novel Method in Dosimetry and Radon Measurement
Introduction Radon is a colorless and tasteless gas which exists in most soils. It is a substance that poses a potential risk for lung cancer in case a person is exposed to high levels over long periods of time. The Environmental Protection Agency (EPA) estimates that 90% of lung cancers per year are caused by radon. The aim of this paper is to estimate the absorbed doses of 222Rn by MCNPX simu...
متن کاملEvaluation of Absorbed Dose of Critical Organ in Rando Phantom under Head, Abdomen and Pelvis Spiral CT Scan by Thermo Luminescent Dosimetery - TLD
Background & Objectives: Computed tomography (CT) represents 11% of all diagnostic radiology procedures but it contributes to almost 67% of the total effective dose to the human population. In head and neck CT which consist of 1/3 of total CT scans, other critical organs such as lenses and thyroid are in the radiation field. Also in the abdomen and pelvis scan, irradiation of ovaries is unavoid...