conventional and spiral ct dose indices in yazd general hospitals, iran

نویسندگان

f. bouzarjomehri department of medical physics, shahid sadoghi university of medical sciences, yazd, iran

m. h. zare department of medical physics, isfahan university of medical sciences, isfahan, iran

d. shahbazi department of medical physics, isfahan university of medical sciences, isfahan, iran

چکیده

abstract   background: while the benefits of computed tomography (ct) are well known in accurate diagnosis, those benefits are not risk free. ct is a device with higher patient dose in comparison with other conventional radiological procedures. is the reduction of exposures by requiring optimization of ct procedures [a principle concern in radiological protection]? in this study, the radiation dose of conventional and spiral ct were investigated and compared with european commission reference dose levels (ec rdls).   materials and methods: the dosimetric quantities proposed in the european guidelines (eg) for ct are weighted computed tomography dose index (ctdiw) for a single slice for axial scanning or per rotation for helical scanning and dose-length product (dlp) for a complete examination. the patient-related data were collected for brain, neck, chest, abdomen and pelvis examination s in each scanner. for each type of examination, 10 typical patients were randomly included. ctdi with an active length of 10cm was measured in two ct scanners by using unfors (mult-o-meter 601) in head and body phantom (pmma) with 16 cm and 32 cm in diameter respectively. mean values of ctdiw , dlp and effective dose(ed) were estimated for those examinations .   results: ctdiw had a range of 15.8-24.7 mgy for brain, 16.1-30.6mgy for neck, 6.8-9.2 mgy for chest, 6.8-9.8 mgy for abdomen and pelvis. dlp had a range of 246.4-397.7 mgy.cm for brain, 104.6-262.2 mgy.cm for neck, 135-248.4 mgy.cm for chest, 187-298.9 mgy.cm for abdomen and 197.2-319.4 mgy.cm for pelvis. the mean values of effective dose were 0.74 msv for brain, 0.9 msv for neck, 3.1 msv for chest, 3.7 msv for abdomen and 5 msv for pelvis.   conclusion: the obtained results in this study have shown that ctdiw and dlp are lower than ec rdls and other studies, in other words , the performance of all scanners has been satisfactory as far as ctdiw and dlp are concerned. the ctdiw and dlp in the conventional ct are higher than the spiral ct values . with regard to alara principle, for the establishment of reference dose levels, the radiation dose with spiral ct scanners should be taken into account.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Conventional and spiral CT dose indices in Yazd general hospitals, Iran

  ABSTRACT   Background: While the benefits of Computed Tomography (CT) are well known in accurate diagnosis, those benefits are not risk free. CT is a device with higher patient dose in comparison with other conventional radiological procedures. Is the reduction of exposures by requiring optimization of CT procedures [a principle concern in radiological protection]? In this study, the radiatio...

متن کامل

Patient dose resulting from CT examinations in Yazd, Iran

Background: With the introduction of computed tomography in diagnostic radiology a new and fundamentally different imaging modality has become available. Meanwhile, it is clear that the absorbed doses by the patients during CT were relatively high in comparison with those of other diagnostic radiology techniques. The aim of this survey was to determine the average absorbed dose in Yazd provin...

متن کامل

Hospitals’ Readiness for Clinical Governance Implementation in Educational Hospitals of Yazd, Iran

BACKGROUND Clinical governance is a systematic approach to maintaining and improving the quality of patient care. This study aimed to assess some Iranian educational hospitals' readiness for clinical governance implementation through the organizational climate. METHODS It was a cross-sectional study that used the Clinical Governance Climate Questionnaire (CGCQ) in three educational hospitals ...

متن کامل

patient dose resulting from ct examinations in yazd, iran

background: with the introduction of computed tomography in diagnostic radiology a new and fundamentally different imaging modality has become available. meanwhile, it is clear that the absorbed doses by the patients during ct were relatively high in comparison with those of other diagnostic radiology techniques. the aim of this survey was to determine the average absorbed dose in yazd province...

متن کامل

Breast dose in conventional and digital mamography

Introduction: The amount of doses received in mammography is one of the biggest concerns in the world. According to some statistics, every 12 minutes a woman dies due to breast cancer. Mammography concerned to detect breast cancer at an early stage of the disease as well as for screening, however due to the risk of exposure, the increasingly and widespread use of mammography fo...

متن کامل

Assessment of Patient Dose from CT Examinations in Khorasan, Iran

Introduction Computed Tomography scans are a very important tool for diagnosis and assessment of response to treatment in the practice of medicine. Ionizing radiation in medical imaging is undoubtedly one of the most powerful diagnostic tools in medicine. Yet, as with all medical interventions, there are potential risks in addition to the clear potential benefits. Materials and Methods Two refe...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید


عنوان ژورنال:
iranian journal of radiation research

جلد ۳، شماره ۴، صفحات ۱۸۳-۱۸۹

کلمات کلیدی

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023