dosimetry limitations and pre-treatment dose profile correction for sliding window imrt

نویسندگان

g. grigorov grand river regional cancer center, grand river hospital, p.o. box 9056, 835 king street west, kitchener, on, canada n2g 1g3

j.c.l. chow radiation medicine program, princess margaret hospital, 610 university avenue, toronto, on, canada m5g 2m9

n. yazdani grand river regional cancer center, grand river hospital, p.o. box 9056, 835 king street west, kitchener, on, canada n2g 1g3

چکیده

background: this work investigated the dosimetry limitations of the random and systematic uncertainties of sliding window (sw) intensity modulated radiation therapy (imrt). materials and methods: a varian 21ex linear accelerator, pinnacle3 treatment planning system and radiographic film dosimetry was used. the limitations of the sw were studied using beam modulation ranging from 2 to 100 mu/beam, dr from 100 to 600 mu min-1, lv from 1 to 5 cm s-1 and field size up to 12 × 12 cm2. the random and systematic errors were investigated using clinical and flat beams, as well as beams of high profile modulation including linear, exponential, and sinusoidal profiles. results: the leading edge and plateau of the sw profiles have a significant deformation for higher dr and for beams of < 10 mus/beam. it was found that the error is directly proportional to the dr and lv, and inversely proportional to the number of mu/beam. conclusion: the high dr and lv are limiting factors, producing random profile deformation when sw beams of small number of mu/beam are delivered. a very good agreement was found between the planned and delivered geometrical and clinical dose profiles when beams > 10 mus irradiated by a dr from 100 to 600 mu min-1 and lv from 1 to 5 cm s-1. after the proposed correction, an average difference < 0.5% for clinical profiles was measured for beams irradiated with dr = 600 mu min-1 and lv= 5 cm s-1. it was concluded that this correction methodology may serve as a pre-treatment quality assurance tool for sw imrt beams. iran. j. radiat. res., 2010 8 (2): 61-74

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Dosimetry limitations and pre-treatment dose profile correction for sliding window IMRT

Background: This work investigated the dosimetry limitations of the random and systematic uncertainties of sliding window (SW) intensity modulated radiation therapy (IMRT). Materials and Methods: A Varian 21EX linear accelerator, Pinnacle3 treatment planning system and radiographic film dosimetry was used. The limitations of the SW were studied using beam modulation ranging from 2 to 10...

متن کامل

Dosimetry limitations and pre-treatment dose profile correction for sliding window IMRT

*Corresponding author: Dr. Grigor Grigorov, Medical Physics Department, Grand River Regional, Cancer Center, PO Box 9056, 835 King St West, Kitchener, ON, Canada N2G 1G3. E-mail: [email protected] Background: This work investigated the dosimetry limitations of the random and systematic uncertainties of sliding window (SW) intensity modulated radiation therapy (IMRT). Materials and Me...

متن کامل

Clinical consequences of changing the sliding window IMRT dose rate

Changing pulse repetition frequency or dose rate used for IMRT treatments can alter the number of monitor units (MUs) and the time required to deliver a plan. This work was done to develop a practical picture of the magnitude of these changes. We used Varian's Eclipse Treatment Planning System to calculate the number of MUs and beam-on times for a total of 40 different treatment plans across an...

متن کامل

Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification.

The aim of this study was to demonstrate how dosimetry with an amorphous silicon electronic portal imaging device (a-Si EPID) replaced film and ionization chamber measurements for routine pre-treatment dosimetry in our clinic. Furthermore, we described how EPID dosimetry was used to solve a clinical problem. IMRT prostate plans were delivered to a homogeneous slab phantom. EPID transit images w...

متن کامل

Dose calibration of EPIDs for segmented IMRT dosimetry

The purpose of this study was to investigate the dose response of amorphous silicon (a-Si) electronic portal imaging devices (EPIDs) under different acquisi- tion settings for both open jaw defined fields and segmented intensity-modulated radiation therapy (IMRT) fields. Four different EPIDs were used. Two Siemens and one Elekta plus a standalone Perkin Elmer research EPID. Each was operated wi...

متن کامل

Assessment of a 2D EPID-based Dosimetry Algorithm for Pre-treatment and In-vivo Midplane Dose Verification

Introduction: The use of electronic portal imaging devices (EPIDs) is a method for the dosimetric verification of radiotherapy plans both pretreatment and in-vivo. The aim of this study was to test a 2D EPID-based dosimetry algorithm for dose verification of some plans inside a homogenous and anthropomorphic phantom and in-vivo, as well. Materials and Methods: </strong...