standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties
نویسندگان
چکیده
background: absolute dosimetry of external beam radiotherapy is carried out by the use of ionization chambers. these chambers must be calibrated at a standard dosimetry laboratory before any use in clinical dosimetry. the secondary standard dosimetry laboratory of iran (ssdl) has the duty of calibrating the ionization chambers used in radiotherapy centers in iran. materials and methods: the present work has described traceability of ssdl radiation measurement standards to relevant international standards, and calibration of therapy level ionization chambers in terms of air kerma and absorbed dose to water against 60co gamma radiation, as well as uncertainty evaluation of calibration coefficients. results: the expanded uncertainties in the determination of air kerma and absorbed dose to water are estimated to be 2% and 2.3% at approximately 95% confidence level, respectively. conclusion: in order to maintain the requirement of ±5% accuracy in the dose delivery, the combined standard uncertainty of the other factors in the dose delivery i.e., dose measurement set-up, dose calculations, treatment planning, patient set-up, etc, should be less than 2.3%. iran. j. radiat. res., 2010 8 (3): 195-199
منابع مشابه
Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties
Background: Absolute dosimetry of external beam radiotherapy is carried out by the use of ionization chambers. These chambers must be calibrated at a standard dosimetry laboratory before any use in clinical dosimetry. The secondary standard dosimetry laboratory of Iran (SSDL) has the duty of calibrating the ionization chambers used in radiotherapy centers in Iran. Materials and Methods...
متن کاملStandard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties
*Corresponding author: Dr. Mostafa Ghafoori, Radiation Dosimetry and Monitoring group, Agricultural, Medical and Industrial Research School (AMIRS), Nuclear Science and Technology Research Institute (NSTRI), Atomic Energy Organization of Iran (AEOI), P.O. Box 31485-498 Karaj, Iran. Fax: +98 261 4464058 E-mail: [email protected] Background: Absolute dosimetry of external beam radiotherapy is c...
متن کاملStability Study of Ionization Chambers in Standard Mammography Radiation Beams
The ionization chambers are widely utilized in diagnostic radiology quality control and dosimetry. They are simple to use, and can be found commercially in various types depending on their applications. They are the recommended devices for the determination of the beam qualities in mammographic range. To assure the reliability of the measurements, the stability of the response of the ionization...
متن کاملThe calibration and use of plane-parallel ionization chambers for dosimetry of electron beams.
The AAPM TG 39 protocol has proposed three different methods of calibrating plane-parallel ionization chambers, i.e., in-phantom irradiation with a high-energy electron beam and in-phantom and in-air 60Co irradiation. To verify the consistency of the three methods, we have measured Ngaspp values using each of these techniques for the five most commonly used plane-parallel chambers considered by...
متن کاملevaluation of sadr eminence in safavid period
چکیده: یکی از دوره های مهم تاریخی ایران به لحاظ تأمین استقلال ملی مذهبی و حتی تأثیر آن بر فرهنگ و مذهب ایرانیان، دوره صفویه است. رسمیت دادن و رواج مذهب شیعه توسط شاه اسماعیل اول، یکی از مهمترین اقدامات این دولت محسوب می شود. بنابراین برای اجرای این سیاست، وی منصب صدارت را به عنوان منصبی مذهبی- حکو متی ایجاد کرد .این منصب از دوره ی تیموریان ایجاد شده بود ولی در اواخر این دوره اهمیت بیشتری یافت...
15 صفحه اولEvaluation of Correction Factors Applied in Photon Calibration of NIS TE Neutron Ionization Chambers
Calibrations of two tissue equivalent (TE) ionization chamber were made in five photon beams (100 kV, 180 kV, 250 kV, Cs and Co) with two different pure gases namely acetylene C2 H2 and carbon dioxide CO2. The different calibration factors were compared both for in-air and in-water phantom, the measurements were performed according to the international atomic energy agency (IAEA) recommendation...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
iranian journal of radiation researchجلد ۸، شماره ۳، صفحات ۱۹۵-۱۹۹
کلمات کلیدی
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023