background: in conventional radiation therapy, regarding normal tissue tolerance, the treatment of bulk tumors is one of the remaining challenges. grid radiation therapy (grt) is a technique to deliver high doses, approximately 15 – 20 gy per fraction, to several small volumes located in a large radiation field. this can be performed using a grid block. the current work has concentrated on the dosimetric characteristics of a designed mega-voltage grid, used for a unique treatment modality. materials and methods: all measurements performed using a neptune linear accelerator (9 mv photon beam). a square 16 × 16 array grid block was designed and fabricated. several dosimetric characteristics including: depth dose, valley to peak (vtp) ratio, and grid out-put factor were evaluated using a calibrated diode dosimeter for a range of radiation fields. results: the percent depth dose curves, measured in the presence of grid block, lie within those measured for the corresponding open field and a narrow beam. at the dmax, the vtp ratio was found to be within 17% - 28%, while these ranges between 23% - 35% at a depth of 10 cm. the grid out -put factor found to be 0.78 and it slightly decreases with increasing of radiation field size. conclusion: the vtp ratio found to be dependent strongly on the grid design and manufacturing properties. however, other parameters such as radiation field size and the depth of measurement should also be addressed as important factors. the measured dosimetric characteristics of grid block indicate that the mega-voltage grid therapy can be applied as a possible clinical modality for palliative cases. iran. j. radiat. res., 2011 9(1): 49-56
منابع مشابه
Evaluation of dosimetric characteristics of a grid block fabricated for Mega-voltage grid therapy purposes
Background: In conventional radiation therapy, regarding normal tissue tolerance, the treatment of bulk tumors is one of the remaining challenges. Grid Radiation Therapy (GRT) is a technique to deliver high doses, approximately 15 – 20 Gy per fraction, to several small volumes located in a large radiation field. This can be performed using a grid block. The current work has concentrate...
متن کاملeffect of sub-grid scales on large eddy simulation of particle deposition in a turbulent channel flow
چکیده ندارد.
15 صفحه اولassessment of the effect of honey as a topical therapy for intra oral wound healing in rat.
چکیده ندارد.
15 صفحه اولEvaluation of Reliability Parameters in Micro-grid
Evaluation of the reliability parameters in micro-grids based on renewable energy sources is one of the main problems that are investigated in this paper. Renewable energy sources such as solar and wind energy, battery as an energy storage system and fuel cell as a backup system are used to provide power to the electrical loads of the micro-grid. Loads in the micro-grid consist of interruptible...
متن کاملGrid Characteristics and Uses: A Grid Definition
This paper discusses the concept of grid towards achieving a complete definition using main grid characteristics and uses found in literature. Ten definitions extracted from main literature sources have been studied allowing the extraction of grid characteristics while grid uses are defined in terms of the different types of application support provided by grids. A grid definition is proposed u...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
iranian journal of radiation researchجلد ۹، شماره ۱، صفحات ۴۹-۵۶
کلمات کلیدی
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023