monte carlo calculation of shielded colpostat effect on rectum received dose in high dose rate brachytherapy with cobalt-60 sources

نویسندگان

a. mesbahi medical physics department, medical school, tabriz university of medical sciences, tabriz, iran

a. haghzadeh engineering faculty, medical radiation engineering department, azad university, science and research branch, tehran, iran

a.r. naseri radiation oncology department, imam hospital, tabriz, iran

a.r. naseri engineering faculty, medical radiation engineering department, azad university, science and research branch, tehran, iran

چکیده

background: in the current study the effect of shielded colpostat on rectum received dose was calculated in cervical brachytherapy using monte carlo (mc) method. materials and methods: two 60co sources of gzp6 brachytherapy unit used for intracavitary treatments were simulated using mcnpx monte carlo code. also the two types of colpostats including shielded and unshielded were simulated inside a water phantom. the radial dose function, depth doses and dose distribution around sources were calculated with spatial dose resolution of 2×2×2 mm. the effect of tungsten shields on rectum dose was studied. moreover, the accuracy of treatment planning system (tps) was verified comparing to mc results. results: our results showed that shielded colpostats reduce the rectum dose up to 7% according to the mc calculations. also, it was found less than 3% difference between tps and mc results in percentage depth dose and radial doses for unshielded case. conclusion: it can be concluded that the application of shielded colpostats provide a relatively mild protection (7% reduction in dose) for the rectum in intracavitary brachytherapy due to the higher energy of 60co photons. however, application of shielded colpostats in treatment planning system is desirable for more accurate treatments.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Monte Carlo calculation of shielded colpostat effect on rectum received dose in high dose rate brachytherapy with Cobalt-60 sources

Background: In the current study the effect of shielded colpostat on rectum received dose was calculated in cervical brachytherapy using Monte Carlo (MC) method. Materials and Methods: Two 60Co sources of GZP6 brachytherapy unit used for intracavitary treatments were simulated using MCNPX Monte Carlo code. Also the two types of colpostats including shielded and unshielded were simulated inside ...

متن کامل

Applicator Attenuation Effect on Dose Calculations of Esophageal High-Dose Rate Brachytherapy Using EDR2 Film

Introduction Interaluminal brachytherapy is one of the important methods of esophageal cancer treatment. The effect of applicator attenuation is not considered in dose calculation method released by AAPM-TG43. In this study, the effect of High-Dose Rate (HDR) brachytherapy esophageal applicator on dose distribution was surveyed in HDR brachytherapy. Materials and Methods A cylindrical PMMA phan...

متن کامل

Radial dose functions of GZP6 intracavitary brachytherapy 60Co sources: treatment planning system versus Monte Carlo calculations

Background: The Monte Carlo (MC) method is not only used for dose calculations around brachytherapy sources but also for benchmarking treatment planning systems (TPS) calculations. Materials and Methods: Three 60Co sources of GZP6 brachytherapy unit were simulated using MCNP4C MC Code. The radial dose functions were calculated by MC method and GZP6 TPS and were compared. Results: There was a go...

متن کامل

New formula for Calculation of Cobalt-60 Percent Depth Dose

Introduction:  On  the  basis  of  percent  depth  dose  (PDD)  calculation,  the  application  of  dosimetry  in  radiotherapy  has  an  important  role  to  play  in  reducing  the  chance  of  tumor  recurrence. The aim of this study is to introduce a new formula for calculating the centeral axis  percent depth doses of Cobalt-60 beam.   Materials and Methods: In the present study, based on ...

متن کامل

A Monte Carlo Study on the Effect of Various Neutron Capturers on Dose Distribution in Brachytherapy with 252Cf Source

Background: In neutron interaction with matter and reduction of neutron energy due to multiple scatterings to the thermal energy range, increasing the probability of thermal neutron capture by neutron captures makes dose enhancement in the tumors loaded with these materials. Objective: The purpose of this study is to evaluate dose distribution in the presence of 10B, 157Gd and 33S neutron...

متن کامل

Dosimetric investigation of high dose rate Ir-192 source with Monte Carlo method

Background: This study aims to calculate the air-kerma strength (SK), the dose rate constant (∧) and the dose rate profiles of a Gammamed 12i Ir-192 source by using the Monte Carlo technique and to compare the dose rate values with those calculated by the Abacus HDR treatment planning system (TPS). Materials and Methods: Air-kerma strength (in units of U; 1 U= μGy m2 h-1) and the dose ra...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید


عنوان ژورنال:
iranian journal of radiation research

جلد ۱۳، شماره ۲، صفحات ۱۶۵-۱۷۱

کلمات کلیدی
[ ' i n t r a c a v i t a r y b r a c h y t h e r a p y ' , ' s h i e l d e d c o l p o s t a t ' , ' r e c t u m d o s e ' , ' h i g h d o s e r a t e b r a c h y t h e r p y ' , ' c o b a l t ' , 6 0 ]

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023