the accuracy of various torque wrenches used in dental implant systems
نویسندگان
چکیده
introduction: the purpose of this study was to compare the accuracy of the torque wrenches used in different dental implant systems. methods: we evaluated 42 torque wrenches used in different dental clinics in mashhad, iran, using a digital torque meter (mark 10). high (25, 30 and 35 n·cm) and low (15 n·cm) levels of torque were examined. ten tests were performed on each wrench, and the mean value was considered as the real torque of the instrument. different characteristics (model (spring or friction), system, duration of use, sterilization, calibration) of each wrench were also recorded. the difference between the torque applied by the instrument and the target torque required was calculated numerically and as a percentage. a one-way anova and student's t-test were used for statistical analysis. results: there was a significant difference between the error at higher torques in the spring wrenches compared with the friction wrenches (p< 0.05). at higher torques, an error greater than 10% was more common in the friction wrenches (29.4%) than in the spring wrenches (4.3%). no significant differences were observed regarding the duration instruments usage and the mean numerical error at high and low torque. in the wrenches that had been used for more than three years, 21.1% of samples showed an error of more than 10%, compared with 9.5% in wrenches that had been used for less than three years (p=0.39). at higher torques the straumann system produced the least error and the biohorizon system produced the greatest error which was significantly greater than the other systems (p< 0.05). conclusion: our results indicate that spring wrenches produce more accurate results than friction wrenches; however, friction wrenches are more reliable at lower torques than higher torques. the length of time in use and sterilization of torque wrenches does not affect the function of the instruments significantly. the precision of the instrument system used is also important.
منابع مشابه
The accuracy of various torque wrenches used in dental implant systems
Introduction: The purpose of this study was to compare the accuracy of the torque wrenches used in different dental implant systems. Methods: We evaluated 42 torque wrenches used in different dental clinics in Mashhad, Iran, using a digital torque meter (Mark 10). High (25, 30 and 35 N·cm) and low (15 N·cm) levels of torque were examined. Ten tests were performed on each wrench, and the mean va...
متن کاملThe Accuracy of New and Aged Mechanical Torque Devices Employed in Five Dental Implant Systems
Purpose Friction-style and spring-style torque wrenches are used to tighten implant abutments and prosthetic screws. The mechanical stability of these torque wrenches is crucial for the implant-abutment connection. The purposes of this study were to assess the performance of five brands (Straumann, Zimmer, Implant KA, Bredent, and Biohorizons) of wrench and to evaluate possible changes in appli...
متن کاملManual maximum closing torque in three implant abutment systems
Background and Aims: Diferences whithin shape and size of hand wrenches in diferent systems could affect the torque applied. Therefore, evaluation of wrenches has been emphasized. The aim of the present study was to compare the torque values of the long and short torque wrenches in different implant systems including ITI, Implantium and Nobel Biocare. Materials and Methods: In this experimenta...
متن کاملIntended and Achieved Torque of Implant Abutment's Screw using Manual Wrenches in Simulated Clinical Setting.
OBJECTIVE To measure the difference between the intended torque and the achieved torque by the operator using the spring-style mechanical torque-limiting device (MTLD). MATERIALS AND METHODS Inexperienced and experienced clinicians used one spring-type MTLD to torque two abutment screws of each anterior and posterior implants, which were attached to two digital torque meters through a jaw mod...
متن کاملThe Frequency of Dentists’ Use and Satisfaction of Different Dental Implant Systems in Qom City, (Iran)
Abstract Background and Objectives: Considering the increasing use of implant, many dentists are willing to use implants with the highest quality and ease of work. Therefore, the present study was performed to investigate the frequency of use and satisfaction of the dentists from different dental implant systems in Qom city. Methods: In this descriptive and cross-sectional study, a questionn...
متن کاملEvaluation of Panoramic Radiography Accuracy in Estimating the Crestal Width of Dental Implant in Different Jaw Regions
Introduction: The aim of this study was to determine the accuracy of digital panoramic radiography in estimating the crestal width of implant fixtures after surgery. Materials and Methods:We selected 241 digital panoramic images after implant insertion which had documented information of their actual size. The crestal width was measured on radiographs. We compared two measuring methods in diff...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
journal of dental materials and techniquesجلد ۲، شماره ۲، صفحات ۳۸-۴۴
کلمات کلیدی
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023