cyclic fatigue resistance and fractographic analysis of race and protaper rotary niti instruments
introduction: this study investigated the fatigue and fracture modes of race and protaper rotary instruments. materials & methods: fatigue resistance was evaluated by rotating the files 30 ° or 60 ° and with 2 or 5mm radius of curvature. race taper 06 size 25 and protaper f1 files (n=40) were used. the number of rotations to failure was analyzed by two-way anova and independent sample t-tes...
متن کاملCyclic Fatigue Resistance and Fractographic Analysis of Race and Protaper Rotary NiTi Instruments
INTRODUCTION This study investigated the fatigue and fracture modes of RaCe and ProTaper rotary instruments. MATERIALS AND METHODS Fatigue resistance was evaluated by rotating the files 30° or 60° and with 2 or 5mm radius of curvature. RaCe taper 06 size 25 and ProTaper F1 files (n=40) were used. The number of rotations to failure was analyzed by two-way ANOVA and independent sample t-test. F...
متن کاملCyclic fatigue resistance of rotary NiTi instruments after simulated clinical use in curved root canals.
The aim of this study was to assess cyclic fatigue resistance in rotary nickel-titanium instruments after simulated clinical use in curved root canals (40-degree, 5-mm radius curve). Thirty-six RaCe rotary NiTi files, size #5, taper 0.04, were divided into 3 groups: Groups A, B and C with one, three and five cycles of use, respectively. Time to failure was recorded with a stopwatch in seconds a...
متن کاملCyclic fatigue of NiTi rotary instruments in a simulated apical abrupt curvature.
AIM To evaluate the cyclic fatigue resistance of five NiTi rotary systems in an abrupt apical curvature. METHODOLOGY Cyclic fatigue testing was performed in stainless-steel artificial canals with a 2-mm radius of curvature and an angle of curvature of 90 degrees constructed to the dimensions of the instruments tested. The middle of the simulated curvature was 2.5 mm from the tip of the instru...
متن کاملCyclic fatigue resistance and three-dimensional analysis of instruments from two nickel-titanium rotary systems.
AIM To determine how instrument design affects the fatigue life of two nickel-titanium (Ni-Ti) rotary systems (Mtwo and ProTaper) under cyclic fatigue stress in simulated root canals. METHODOLOGY Cyclic fatigue testing of instruments was performed in stainless steel artificial canals with radii of curvature of 2 or 5 mm and an angle of curvature of 60 degrees . A total of 260 instruments were...