Verification of UriSed 3 PRO automated urine microscope in regional laboratory environment
نویسندگان
چکیده
Ten UriSed 3 PRO automated microscopes (77 Elektronika, Hungary) were verified for nine HUSLAB laboratories with 160 000 annual urine samples. Particle counting of the primary instrument was against reference visual microscopy 463 specimens, and culture on chromogenic agar plates parallel 396 specimens. Nine secondary instruments compared pairwise instrument. Relative imprecisions to Poisson distribution, R(CV), estimated be 1.0 white blood cell (WBC) 1.5 red (RBC) counts, respectively. Spearman’s correlations rS = 0.94 WBC, 0.87 RBC, 0.82 squamous epithelial (SEC) counts. Agreement (Cohen’s weighted kappa) 0.89 0.88 SEC, 0.59 combined casts, 0.49 non-squamous cells (NEC). Bacteria detected a sensitivity 90% specificity 39 at 107 CFB/L (104 CFU/mL). Created flagging limits allowed reporting 70–75% patient results. adopted into routine use after acceptance verification.
منابع مشابه
Diagnostic accuracy of UriSed automated urine microscopic sediment analyzer and dipstick parameters in predicting urine culture test results
INTRODUCTION Urinary tract infection (UTI) is one of the most common types of infection. Currently, diagnosis is primarily based on microbiologic culture, which is time- and labor-consuming. The aim of this study was to assess the diagnostic accuracy of urinalysis results from UriSed (77 Electronica, Budapest, Hungary), an automated microscopic image-based sediment analyzer, in predicting posit...
متن کاملA Fully Automated Environment for Verification of Virtual Prototypes
The extremely dynamic and competitive nature of the wireless communication systems market demands ever shorter times to market for new products. Virtual prototyping has emerged as one of the most promising techniques to offer the required time savings and resulting increases in design efficiency. A fully automated environment for development of virtual prototypes is presented here, offering max...
متن کاملAutomated 3-D Reconstruction Using a Scanning Electron Microscope
Methods for both the accurate calibration and also the use of an Environmental Scanning Electron Microscope (ESEM) for accurate 3D reconstruction are described. Unlike previous work, the proposed methodology does not require either known motions of a target or use of calibration target with accurate fiducial marks. Experimental results from the calibration studies show the necessity for taking ...
متن کاملcellular behavior and scanning electron microscope evaluation of pro-root mta, root mta and modified mta on fibroblast l929
چکیده ندارد.
15 صفحه اولAutomated Flow Cytometry: An Alternative to Urine Culture in a Routine Clinical Microbiology Laboratory?
The urine culture is the "gold standard" for the diagnosis of urinary tract infections (UTI) but constitutes a significant workload in the routine clinical laboratory. Due to the high percentage of negative results, there is a need for an efficient screening method, with a high negative predictive value (NPV) that could reduce the number of unnecessary culture tests. With the purpose of improvi...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ژورنال
عنوان ژورنال: Clinica Chimica Acta
سال: 2021
ISSN: ['1873-3492', '0009-8981']
DOI: https://doi.org/10.1016/j.cca.2021.01.005