VALIDATION OF WARFARIN ANALYSIS METHOD IN HUMAN BLOOD PLASMA USING HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY WITH FLUORESCENCE DETECTION
نویسندگان
چکیده
Objective: Validation of analysis method is important, especially in analyzing narrow-index drugs such as warfarin (WF). This study aimed to obtain a validated human plasma according European Medicine Agency guidelines.
 Methods: The optimum conditions for the using fluorescence detector HPLC with Chiralcel OD-RH column (4.6 x 150 mm i.d., 5μm); guard (4.0 10 mm, 5μm), temperature 45 °C. mobile phase used was acetonitrile: phosphate buffer pH 2 (40:60), an isocratic flow rate 1 ml/min and injection volume 20 μl. Excitation emission wavelengths were 310 350 nm (warfarin) 300 400 (griseofulvin). retention time griseofulvin 6-7.5 min; R-warfarin 10-11.5 S-warfarin 14-16 min.
 Results: result this validation obtained condition. yielded LOD values 0.0674 ppm (R-warfarin) 0.0897 (S-warfarin). LOQ 0.225 0.298 Linearity at concentrations 0.2-3 line equation y = 0.0705x+0.0704 R2 0.978 0.0513x+0.0297 R² 0.9924 S-warfarin. 75% seven met reverse concentration requirements, which below±15%. requirements accuracy precision within between runs, selectivity carryover where %RSD %diff below±15%
 Conclusion: analytical can be declared valid sample measurement pharmacokinetic studies.
منابع مشابه
A quick and easy high performance liquid chromatography method for evaluation of cefixime in human plasma
Cefixime is essential member of orally energetic third generation cephalosporin and has tremendous activity aligned with many pathogens. The virtual bioavailability of a newly industrial dispersible tablet as compared with a recognized identified formulation. A simple and available reversed-phase HPLC method with UV detection has been urbanized and validate for cefixime evaluate in human p...
متن کاملDetermination of amlodipine in human plasma by high-performance liquid chromatography with fluorescence detection.
A sensitive and specific HPLC method has been developed for the assay of amlodipine in human plasma. The assay involves derivatization with 4-chloro-7-nitrobenzofurazan (NBD-Cl), solid-phase extraction on a silica column and isocratic reversed-phase chromatography with fluorescence detection. Nortriptyline hydrochloride was used as an internal standard. The assay was linear over the concentrati...
متن کاملDetermination of codeine in human plasma by high-performance liquid chromatography with fluorescence detection.
A rapid, reliable and rugged assay for determining codeine in human plasma using reversed-phase high-performance liquid chromatography with fluorescence detection was developed. This analytical method utilized an ion-exchange/mixed-mode solid-phase extraction procedure. The chromatographic separation was achieved using a 150 x 4.6 mm I.D., 3-microns reversed-phase C8 (deactivated for basic anal...
متن کاملHigh-Performance Liquid Chromatography Determination of Methotrexate in Plasma
This article describes a simple and fast high-performance liquid chromatography method for the determination of methotrexate (MTX) in plasma. Samples were collocted from children receiving high-dose MTX at shafa Hospital (Ahvaz University of Medical sciences, Ahvaz, iran) at various times after the end of eachinfusion. Plasma was deproteinized with trichloroacetic acid and the supernatant was i...
متن کاملHigh-Performance Liquid Chromatography Determination of Methotrexate in Plasma
This article describes a simple and fast high-performance liquid chromatography method for the determination of methotrexate (MTX) in plasma. Samples were collocted from children receiving high-dose MTX at shafa Hospital (Ahvaz University of Medical sciences, Ahvaz, iran) at various times after the end of eachinfusion. Plasma was deproteinized with trichloroacetic acid and the supernatant was i...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ژورنال
عنوان ژورنال: International Journal of Applied Pharmaceutics
سال: 2022
ISSN: ['0975-7058']
DOI: https://doi.org/10.22159/ijap.2022.v14s5.27