Validation of a prototype DiodeAir for small field dosimetry
نویسندگان
چکیده
منابع مشابه
Validation of a prototype DiodeAir for small field dosimetry.
Standard commercial diode detectors over-respond within small radiation fields, an effect largely attributable to the relatively high mass-density of silicon. However, Monte Carlo studies can be used to optimise dosimeter designs and have demonstrated that 'mass-density compensation'-for example, introducing a low-density air-gap upstream of a diode's high-density silicon volume-can substantial...
متن کاملEnergy and field size dependence of a silicon diode designed for small-field dosimetry
Purpose: To investigate the energy dependence/spectral sensitivity of silicon diodes designed for small-field dosimetry and obtain response factors (RFs) for arbitrary photon spectra using Monte Carlo (MC) simulations. Methods: The EGSnrc user-code DOSRZnrc was used to calculate the dose deposition in water and in the active volume of a stereotactic diode f...
متن کاملDevelopment of Prototype Iranian male pelvic phantom for internal dosimetry
Introduction: Existing phantoms have been constructed based on Caucasian, non-Caucasian and race-specific datasets. According to previous studies made efforts to present Korean- specific phantoms and Chinese female phantom based on CVH dataset due to compare the resulting internal dosimetry with the Caucasian based data showed possible racial difference in human anatomy between ...
متن کاملMonte carlo simulation of varian clinac iX 10 MV photon beam for small field dosimetry
Background: The lack of lateral electronic disequilibrium (LED) becomes a main problem in small field. This factor affects the dose in target volume cannot predict correctly. In addition, utilization of high-energy linear accelerator (10 MV) can emit some unwanted particles (electron contamination). Therefore, the aim of this study was to characterize head linear accelerator (linac) Varian Clin...
متن کاملReference dosimetry and small - field dosimetry in external beam radiotherapy : Results from a Danish intercomparison study
max. 2000 char.): A comparison of dosimetry methods at different clinics can be used as a means to uncover systematic uncertainties in radiotherapy. To assess the current status of reference dosimetry and small-field dosimetry in clinical practice, a collaborative comparison study involving several dosimetry methods was performed by DTU Nutech at six Danish clinics. The first part of the interc...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ژورنال
عنوان ژورنال: Physics in Medicine and Biology
سال: 2015
ISSN: 0031-9155,1361-6560