Tono-Pen XLR calibration curves for cats, cows and sheep
نویسندگان
چکیده
منابع مشابه
® calibration curves for cats, cows and sheep
The objective of this study was to provide calibration curves for correcting intraocular pressure (IOP) measurements obtained using the Tono-Pen XL ® tonometer in cats, cows and sheep. Twelve eyes from 9 cats, 13 eyes from 7 cows, 10 eyes from 5 sheep were used. The anterior chamber of the eye was cannulated in vivo , in situ (immediately post mortem) or ex vivo with a fine needle and IOP was v...
متن کاملTono-Pen tonometry in normal and in post-keratoplasty eyes.
Oculab Tono-Pen tonometry was compared with Goldmann applanation tonometry in 82 eyes of 82 patients with normal corneas and in 54 eyes of 54 patients who had undergone penetrating keratoplasty and whose corneas did not preclude the use of Goldmann tonometer. We found that the intraocular pressure (IOP) in 48% of the eyes with normal corneas and in 57% after keratoplasty has different measureme...
متن کاملNoninvasive measurement of rat intraocular pressure with the Tono-Pen.
PURPOSE The purpose of this study was to evaluate the Tono-Pen 2 tonometer for measuring intraocular pressure (IOP) in the living rat eye. METHODS One eye from each of 20 adult, anesthetized brown Norway rats (group 1) was cannulated and simultaneously connected to a syringe and a pressure transducer with a chart recorder. We increased IOP from 15 to 45 mmHg in 5-mmHg increments and obtained ...
متن کاملIntraocular pressure measurement in cynomolgus monkeys. Tono-Pen versus manometry.
PURPOSE In living cynomolgus monkey eyes, the authors compared manometrically set and measured intraocular pressure (IOP) with simultaneous IOP readings obtained with the Tono-Pen (TP), a handheld applanation tonometer based on the Mackay-Marg principle. METHODS In three pentobarbital-anesthetized cynomolgus monkeys, IOP was set and measured manometrically after anterior chamber cannulation t...
متن کاملClinical comparison of the ProTon and Tono-Pen tonometers with the Goldmann applanation tonometer.
A clinical evaluation of a new electron ProTon tonometer was performed comparing the values of intraocular pressure (IOP) measured using this instrument with those determined by a similar instrument, Tono-Pen XL, and by Goldmann applanation tonometry. The mean IOP measured in 106 eyes with the ProTon tonometer was not significantly different from that determined with Goldmann applanation, while...