Restenosis Investigation of Two-Stent Placement in the Artery Bifurcation with Different Stenting Techniques Using Computational Fluid Dynamics Analysis

نویسندگان

چکیده

The in-stent restenosis may still occur in patients with a coronary bifurcation stenting. purpose of this study is to investigate the effect relationships between angles artery and three types stenting techniques (T-stenting, Culotte, Kissing techniques), for comparing two kinds arterial shapes using computational fluid dynamics (CFD) analysis. finite element models arteries were reconstructed stereolithography (STL) data an artificial from reverse engineering laser scanning, then each was further modified add angles, including 80°, 113°, 142°. low TAWSS mainly concentrated at LCX branch; more essential discovery that area ration proportion angle. smallest T-stenting detected A B, respectively. Both Culotte always showed higher OSI than T-stenting, high meaningfully located branch. CFD concluded large angle long stent side branch technique has risk restenosis.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Overlap stenting for in-stent restenosis after carotid artery stenting

Our aim was to assess the clinical safety and efficacy of overlap stenting for in-stent restenosis after carotid artery stenting. The study was conducted between July 2008 and February 2015. A database of consecutive carotid artery stenting procedures was retrospectively assessed to identify the cases of in-stent restenosis that were treated with overlap stenting under proximal or distal protec...

متن کامل

Computational fluid dynamics analysis of tandem carotid artery stenoses: Investigation of neurological complications after carotid artery stenting.

BACKGROUND Combined extra- and intracranial carotid artery stenoses, particularly involving multiple lesions, show complex hemodynamic properties and represent a therapeutic dilemma. We used computational fluid dynamics (CFD) to investigate whether insufficient cerebral blood flow (CBF) in a 70-year-old man with tandem stenoses was the cause of aphasia and right hemiparesis after carotid artery...

متن کامل

In Stent Restenosis Predictors after Carotid Artery Stenting

Purpose. The long-term efficacy of carotid artery stenting is debated. Predictors of stent restenosis are not fully investigated. Our aim was to assess the incidence of long term restenosis after CAS and to identify some predictors of restenosis. Methods. We retrospectively selected 189 treated patients and we obtained the survival Kaplan-Meier curves for overall survival, for freedom from stro...

متن کامل

Stent strut fracture-induced restenosis in a bifurcation lesion treated with the crush stenting technique.

Percutaneous treatment of a bifurcation lesion still shows a significant complication rate, mainly because of restenosis at the ostial site of the side branch vessel. Different techniques, such as V-stenting, culottes-stenting or crush stenting, allow full ostial coverage and may therefore achieve uniform drug distribution within the lesion. The crush technique results in a strong mechanical co...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ژورنال

عنوان ژورنال: Applied sciences

سال: 2023

ISSN: ['2076-3417']

DOI: https://doi.org/10.3390/app13084694