Response Letter to Editor In Reference to Letter to Editor about "Adverse Clinical Outcomes in a Primary Modular Neck/Stem System".
We report our experience with 215 recalled neck modular stems due to corrosion. Among the 195 hips with 2 years follow-up, 56% had no clinical symptoms, 26% had groin pain (typical of corrosion), and 17% had other symptoms. Cobalt levels were comparable between asymptomatic (3.4 μg/L, range 0.7-7.3 μg/L) and symptomatic patients (4.0 μg/L range 0-13.2 μg/L). Abnormal imaging findings were seen ...