Quantification and reduction of peripheral dose from leakage radiation on Siemens primus accelerators in electron therapy mode

In this work, leakage radiation from EA200 series electron applicators on Siemens Primus accelerators is quantified, and its penetration ability in water and/or the shielding material Xenolite-NL established. Initially, measurement of leakage from 10 x 10 - 25 x 25 cm2 applicators was performed as a function of height along applicator and of lateral distance from applicator body. Relative to ce...

متن کامل

Radiation leakage dose from Elekta electron collimation system

This study provided baseline data required for a greater project, whose objective was to design a new Elekta electron collimation system having significantly lighter electron applicators with equally low out-of field leakage dose. Specifically, off-axis dose profiles for the electron collimation system of our uniquely configured Elekta Infinity accelerator with the MLCi2 treatment head were mea...

متن کامل

The Quality Control of Intensity Modulated Radiation Therapy (IMRT) for ONCOR Siemens Linear Accelerators Using Film Dosimetry

Introduction Intensity Modulated Radiation Therapy (IMRT) has made a significant progress in radiation therapy centers in recent years. In this method, each radiation beam is divided into many subfields that create a field with a modulated intensity. Considering the complexity of this method, the quality control for IMRT is a topic of interest for researchers. This article is about the various ...

متن کامل

Leakage reduction for the Siemens ModuLeaf

The Moduleaf, an add-on miniature multileaf collimator (MMLC) for the Siemens Oncor linear accelerator, provides high resolution field shaping with a maximum inter-leaf leakage dose of 1.50% at 6 MV. However, beyond the maximum treatment field size, the distribution of leakage and scatter along the y axis is different from that of the x axis, with maximum leakage values of 1.53% and 0.39%, resp...

متن کامل

Neutron Radiation from Medical Electron Accelerators*

Electron accelerators operating above about 10 MeV produce significant quantities of photoneutrons. This is not a new problem and was first reported by Laughlin1 in 1951 for a 23 MeV betatron. In recent years it has drawn considerable attention because of the increased use of higher-energy electron linear accelerators for radiation therapy and also because of the increased awareness of radiatio...