Quality control of VMAT synchronization using portal imaging
نویسندگان
چکیده
منابع مشابه
Quality control of VMAT synchronization using portal imaging
For accurate delivery of volumetric-modulated arc therapy (VMAT), the gantry position should be synchronized with the multileaf collimator (MLC) leaf positions and the dose rate. This study, therefore, aims to implement quality control (QC) of VMAT synchronization, with as few arcs as possible and with minimal data handling time, using portal imaging. A steel bar of diameter 12 mm is accurately...
متن کاملPretreatment quality control of single isocenter half- beam treatment planning technique using an amorphous silicon electronic portal-imaging device (EPID)
Introduction: Electronic portal imaging devices (EPIDs) are fundamentally used for instantaneous verification of the patient set‐up, block shape, and leaf positions during radiation therapy. In radiotherapy, situations arise in which an inclined PTV must be treated mutually with adjacent nodal regions. This methodology is most widely used for matching tangential/lateral breas...
متن کاملTwin machines validation for VMAT treatments using electronic portal-imaging device: a multicenter study
PURPOSE To verify the accuracy of volumetric arc therapy (VMAT) using the RapidArc™ device when switching patients from one single linear accelerator (linac) to a paired energy and mechanics "twin" linac without reoptimization of the original treatment plan. PATIENTS AND METHODS Four centers using 8 linacs were involved in this study. Seventy-four patients previously treated with the 6MV phot...
متن کاملDaily dosimetric quality control of the MM50 Racetrack Microtron using an electronic portal imaging device.
The MM50 Racetrack Microtron, suited for advanced three-dimensional conformal radiotherapy techniques, is a complex machine in various respects. Therefore, for a number of gantry angles, daily quality control of the absolute output and fluence profiles of the scanned beams are mandatory. For the applied photon beams, a fast method for these daily checks, based on dosimetric measurements with th...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ژورنال
عنوان ژورنال: Journal of Applied Clinical Medical Physics
سال: 2015
ISSN: 1526-9914
DOI: 10.1120/jacmp.v16i1.5238