Quality Control of Conventional X-Ray Tube in Three Tertiary Hospitals in South-East, Nigeria
نویسندگان
چکیده

 Background:Quality control of conventional x-ray tube ensures that the patient integral radiation dose is minimized and image quality improved, by controlling beam to reduce scatter radiation.
 Objective:To assess tubes for half value layer (HVL), field light congruence using known standard.
 Methodology:The HVLs were measured calibrated, non-invasive, digital multifunctional detector meter incorporate computer output. The was positioned at center collimated axis with focus-to-image distance (FID) 100 cm. Tube potentials 80 kVp selected, used make exposures. corresponding then recorded. Also, placing 18 cm x 24 cassette loaded film FID collimator in ON position metal markers delineate periphery field. Misalignment calculated from developed radiographs standard formula.
 Results:The HVL ranged 3.40 4.4mmAl. sum misalignment both orthogonal directions 7.21 9.70 % all three centers.
 Conclusion:The within limit centers studied. However, grossly misaligned.
منابع مشابه
Quality assessment of conventional X-ray diagnostic equipment by measuring X-ray exposure and tube output parameters in Great Khorasan Province, Iran
Introduction: Regular implementation of quality control (QC) program in diagnostic X-ray facilities may affect both image quality and patient radiation dose due to the changes in exposure parameters. Therefore, this study aimed to investigate the status of randomly selected conventional radiographic X-ray devices installed in radiology centers of Great Khorasan Province, Iran, to produce the da...
متن کاملImpact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals
ABSTRACT Background: The X-ray machines used for radiodiagnosis should meet certain quality assurance (QA) programmes . These are necessary to have good quality radiographs at reasonably low exposure to patients. Materials and Methods: Dose reduction methods in abdomen X-ray examination were carried out in 10 hospitals in Tehran. This paper presents the work, which was implemented on 200 ...
متن کاملThe Evaluation of Conventional X-ray Exposure Parameters Including Tube Voltage and Exposure Time in Private and Governmental Hospitals of Lorestan Province, Iran
Introduction In radiography, dose and image quality are dependent on radiographic parameters. The problem is caused from incorrect use of radiography equipment and from the radiation exposure to patients much more than required. Therefore, the aim of this study was to implement a quality-control program to detect changes in exposure parameters, which may affect diagnosis or patient radiation do...
متن کاملQuality Control of Conventional Radiology Devices in Selected Hospitals of Khuzestan Province, Iran
Introduction Quality control techniques used to test the components of the radiological system and verify that the equipment is operating satisfactorily. In this study, quality control (QC) assessment of conventional radiology devices was performed in frequently visited radiology centers of Khuzestan province, Iran. Materials and Methods Fifteen conventional radiology devices were examined, bas...
متن کاملEvaluation of X-Ray Radiation Levels in Radiology Departments of Two Educational Hospitals in Ahvaz, Iran
Introduction: The ionizing radiation is increasingly applied in various fields for industrial and medical purposes due to its benefits. The aim of this study was to measure the radiation levels in six radiology departments of two educational hospitals in Ahvaz, Iran. Materials and Methods: The radiation levels were measured at six locations of six radiology departments, including behind the pat...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ژورنال
عنوان ژورنال: Journal of Radiography and Radiation Sciences
سال: 2022
ISSN: ['2736-1071', '2736-1063']
DOI: https://doi.org/10.48153/jrrs.v32i1.223078