A 21-year-old man underwent stent implantation (8Z45CP stent on 18 mm BiB-balloon, Numed, Hopkinton, NY) for a recurrent aortic obstruction, after previous repair of an aortic interruption type A. At the age of two weeks after birth an end-to-side anastomosis of the descending aorta onto the distal arch (left subclavian artery) was performed. Diagnosis of restenosis was based upon upper limb hy...
متن کاملEccentric stent graft compression: an indicator of insecure proximal fixation of aortic stent graft.
PURPOSE The purpose of this study was to determine whether radiographically demonstrated proximal stent graft contour can be used as a marker for security of proximal neck fixation after endovascular aneurysm repair. METHODS Stent graft structure was examined in 100 consecutive patients with abdominal aortic aneurysms who were treated with the stent graft. Stent graft integrity, stent contour...
متن کاملThe Relationship between Temporal Changes in Proximal Neck Angulation and Stent-Graft Migration after Endovascular Abdominal Aortic Aneurysm Repair.
BACKGROUND In recent years, endovascular abdominal aortic aneurysm repair (EVAR) for treating abdominal aortic aneurysms (AAA) has become quite prevalent in Japan. Though little information is available about temporal changes in proximal neck angulation due to the difficulties encountered in measuring the angle. Therefore, we examined temporal changes in proximal neck angulation and its relatio...
متن کاملProximal aortic arch cannulation for proximal ascending aortic aneurysms
INTRODUCTION Different arterial inflow sites have been reported to date for particularly challenging cardiac operations. The ascending aorta, femoral artery, and subclavian artery are the most commonly used sites. Although its use has been reported, the aortic arch has not gained popularity in the performance of cannulation. According to a search performed in the PubMed database, aortic arch ca...
متن کاملDirectional tip control technique for optimal stent graft alignment in angulated proximal aortic landing zones
Angulated anatomy in the aorta, such as tortuous infrarenal aortic necks or steep aortic arches, is a significant challenge for endovascular aortic repair because it often causes inadequate sealing and fixation, which may lead to treatment failure. We have developed a technique using off-the-shelf equipment to precisely control the deployment of stent grafts in challenging landing zones. The ke...