Performance of the HIV Blot 2.2, INNO-LIA HIV I/II Score, and Geenius HIV 1/2 Confirmatory Assay for use in HIV confirmation
نویسندگان
چکیده
منابع مشابه
Evaluation of the Bio-Rad Geenius HIV 1/2 assay as an alternative to the INNO-LIA HIV 1/2 assay for confirmation of HIV infection.
The Bio-Rad Geenius HIV 1/2 assay was evaluated as an alternative to the INNO-LIA HIV 1/2 assay for the confirmation of HIV infection in 198 serum samples reactive to 4th-generation HIV enzyme immunoassays (EIAs). The Geenius assay correctly identified 85% of the samples, compared to 75% identified by the INNO-LIA assay, reduced the number of indeterminate results, and shortened the overall tur...
متن کاملComparison of a conventional HIV 1/2 line immunoassay with a rapid confirmatory HIV 1/2 assay.
The performance of the rapid confirmatory HIV 1/2 assay Geenius was compared with the conventional HIV 1/2 line immunoblot (INNO-LIA HIV I/II Score). One hundred HIV 1/2 confirmed positive samples from donors and patients and 136 negative screening samples from blood donors were evaluated with both assays. A 20 member performance panel consisting of different HIV 1 and 2 subtypes was also analy...
متن کاملEvaluation of the Bio-Rad Geenius HIV-1/2 test as a confirmatory assay
OBJECTIVES We have evaluated the recently Conformité Européenne (CE)-marked Bio-Rad Geenius human immunodeficiency virus (HIV)1/2 as a rapid and simple alternative to western blot for confirmation of HIV screening results. METHODS A total of 160 serum samples were tested: 44 HIV-1 reactive samples by a fourth-generation Murex HIV Ag/Ab and/or Vidas HIV Duo Ultra, five HIV-2 reactive samples, ...
متن کاملOropharyngeal candidiasis in HIV/AIDS patients and non-HIV subjects in the Southeast of Iran
Background and Purpose: Candida species are the common opportunistic pathogens during the course of human immunodeficiency virus (HIV) infection. Oropharyngeal candidiasis (OPC) is generally known as the initial sign of HIV infection. The aim of this study was to compare demographic characteristics and frequency of Candida species between HIV/AIDS patients and non-HIV subjects in Kerman, southe...
متن کاملFalse-positive result of a confirmatory human immunodeficiency virus line immuno assay in an apparently healthy individual--a case report.
A case of a false-positive result of human immunodeficiency virus (HIV) confirmatory immunoblot-based assay is described. Repeatedly borderline reactive anti-HIV screening enzyme immunoassay result obtained in a local hospital resulted in directing the sample to the Slovenian HIV/AIDS Reference Laboratory. In the Reference Laboratory, both anti-HIV screening assays and confirmatory Western blot...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ژورنال
عنوان ژورنال: PLOS ONE
سال: 2018
ISSN: 1932-6203
DOI: 10.1371/journal.pone.0199502