Patient-Specific QA of Spot-Scanning Proton Beams Using Radiochromic Film
نویسندگان
چکیده
منابع مشابه
Patient-Specific QA of Spot-Scanning Proton Beams using Radiochromic Film.
Radiochromic film for spot-scanning QA provides high spatial resolution and efficiency gains from one-shot irradiation for multiple depths. However, calibration can be a tedious procedure which may limit widespread use. Moreover, since there may be an energy dependence, which manifests as a depth dependence, this may require additional measurements for each patient. We present a one-scan protoc...
متن کاملThree‐dimensional gamma criterion for patient‐specific quality assurance of spot scanning proton beams
The purpose of this study was to evaluate the effectiveness of full three-dimensional (3D) gamma algorithm for spot scanning proton fields, also referred to as pencil beam scanning (PBS) fields. The difference between the full 3D gamma algorithm and a simplified two-dimensional (2D) version was presented. Both 3D and 2D gamma algorithms are used for dose evaluations of clinical proton PBS field...
متن کاملReference dosimetry using radiochromic film
The objectives of this study are to identify and quantify factors that influence radiochromic film dose response and to determine whether such films are suitable for reference dosimetry. The influence of several parameters that may introduce systematic dose errors when performing reference dose measurements were investigated. The effect of the film storage temperature was determined by comparin...
متن کاملTowards Effective and Efficient Patient-Specific Quality Assurance for Spot Scanning Proton Therapy
An intensity-modulated proton therapy (IMPT) patient-specific quality assurance (PSQA) program based on measurement alone can be very time consuming due to the highly modulated dose distributions of IMPT fields. Incorporating independent dose calculation and treatment log file analysis could reduce the time required for measurements. In this article, we summarize our effort to develop an effici...
متن کاملModelling of a proton spot scanning system using MCNP6
The aim of this work was to model the characteristics of a clinical proton spot scanning beam using Monte Carlo simulations with the code MCNP6. The proton beam was defined using parameters obtained from beam commissioning at the Skandion Clinic, Uppsala, Sweden. Simulations were evaluated against measurements for proton energies between 60 and 226 MeV with regard to range in water, lateral spo...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ژورنال
عنوان ژورنال: International Journal of Medical Physics, Clinical Engineering and Radiation Oncology
سال: 2017
ISSN: 2168-5436,2168-5444
DOI: 10.4236/ijmpcero.2017.62011