Multicenter cross-calibration of I-123 metaiodobenzylguanidine heart-to-mediastinum ratios to overcome camera-collimator variations
نویسندگان
چکیده
منابع مشابه
Multicenter cross-calibration of I-123 metaiodobenzylguanidine heart-to-mediastinum ratios to overcome camera-collimator variations
BACKGROUND The heart-to-mediastinum ratio (HMR) of (123)I-metaiodobenzylguanidine (MIBG) showed variations among institutions and needs to be standardized among various scinticamera-collimator combinations. METHODS A total of 225 phantom experiments were performed in 84 institutions to calculate cross-calibration coefficients of HMR. Based on phantom studies, a conversion coefficient for each...
متن کاملStandardization of the heart-to-mediastinum ratio of iodine-123-labeled-meta-iodobenzylguanidine uptake using dual energy window method: Feasibility of correction from different camera-collimator combinations
Background. Although heart-to-mediastinum (H/M) ratio in a planner image has been used for practical quantification for I-metaiodobenzylguanidine (MIBG) imaging, standardization of the parameter is not yet established. We hypothesized that the value of H/M can be standardized to the various types of camera-collimator combinations. Methods and Results. Standard phantoms consisting of the heart, ...
متن کاملCross calibration of 123I-meta-iodobenzylguanidine heart-to-mediastinum ratio with D-SPECT planogram and Anger camera
BACKGROUND Cardiac 123I-meta-iodobenzylguanidine (MIBG) uptake is quantified using the heart-to-mediastinum ratio (HMR) with an Anger camera. The relationship between HMR determined using D-SPECT with a cadmium-zinc-telluride detector and an Anger camera is not fully understood. Therefore, the present study aimed to define this relationship using images derived from a phantom and from patients....
متن کاملLetter to the Editor: Interobserver Variability of Heart-to-Mediastinum Ratio in I-123 MIBG Sympathetic Imaging
One of the significant contributors to the variation of H/ M ratio in cardiac imaging, are the settings of the regions of interest (ROIs) that are used. Currently, as correctly described by Chen et al. [1, 2], there are three main approaches ROI is placed along the anterior part of the LV wall [2]. In our current ongoing study in pediatric patients by van der Palen et al.., we compared both app...
متن کاملMyocardial Iodine-123-Metaiodobenzylguanidine (I-MIBG) Imaging in Brugada Syndrome
(I-MIBG) Imaging in Brugada Syndrome To the Editor: We read with interest a recent article in Circulation by Wichter et al1 on cardiac autonomic dysfunction in Brugada syndrome. The authors demonstrated that regionally reduced iodine-123-metaiodobenzylguanidine (I-MIBG) uptake in the inferior and septal left ventricular wall was present in 8 (47%) of 17 patients with Brugada syndrome but in non...