MAGAT gel and EBT2 film-based dosimetry for evaluating source plugging-based treatment plan in Gamma Knife stereotactic radiosurgery
نویسندگان
چکیده
منابع مشابه
MAGAT gel and EBT2 film‐based dosimetry for evaluating source plugging‐based treatment plan in Gamma Knife stereotactic radiosurgery
This work illustrates a procedure to assess the overall accuracy associated with Gamma Knife treatment planning using plugging. The main role of source plugging or blocking is to create dose falloff in the junction between a target and a critical structure. We report the use of MAGAT gel dosimeter for verification of an experimental treatment plan based on plugging. The polymer gel contained in...
متن کاملMRI‐based polymer gel dosimetry for validating plans with multiple matrices in Gamma Knife stereotactic radiosurgery
One of treatment planning techniques with Leksell GammaPlan (LGP) for Gamma Knife stereotactic radiosurgery (GKSRS) uses multiple matrices with multiple dose prescriptions. Computational complexity increases when shots are placed in multiple matrices with different grid sizes. Hence, the experimental validation of LGP calculated dose distributions is needed for those cases. For the current stud...
متن کاملGamma knife stereotactic radiosurgery for radiation-induced meningiomas.
BACKGROUND Radiation-induced meningiomas present a unique clinical dilemma given the fact that patients with these tumors have often received a prior full course of radiotherapy. As such, traditional radiotherapy is limited by lifetime tissue tolerances to radiation, leaving surgery and radiosurgery as attractive treatment options. OBJECTIVES To ascertain the safety and efficacy of Gamma Knif...
متن کاملGamma Knife Stereotactic Radiosurgery for Nonvestibular Cranial Nerve Schwannomas.
BACKGROUND Nonvestibular cranial schwannomas represent a rare type of benign intracranial tumor. Few studies have evaluated the use of stereotactic radiosurgery (SRS) as a primary management option for these lesions. We performed a retrospective review of our institution's experience focusing on efficacy with regard to tumor control and clinical symptom stabilization as well as treatment safety...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ژورنال
عنوان ژورنال: Journal of Applied Clinical Medical Physics
سال: 2012
ISSN: 1526-9914
DOI: 10.1120/jacmp.v13i6.3877