Image quality, diagnostic accuracy, and potential for radiation dose reduction in thoracoabdominal CT, using Sinogram Affirmed Iterative Reconstruction (SAFIRE) technique in a longitudinal study
نویسندگان
چکیده
منابع مشابه
Image quality, diagnostic accuracy, and potential for radiation dose reduction in thoracoabdominal CT, using Sinogram Affirmed Iterative Reconstruction (SAFIRE) technique in a longitudinal study
OBJECTIVE To step-wise evaluate image quality of sinogram-affirmed iterative reconstruction (SAFIRE) in reduced-dose (RD) thoracoabdominal computed tomography (CT) compared to full-dose (FD) and RD filtered back projection (FBP) in a longitudinal study. MATERIALS AND METHODS 122 patients were included in this prospective study. 49 patients (14 men: mean age ± SD, 56±0.4 years; 35 women: 58±1....
متن کاملRadiation Dose Reduction via Sinogram Affirmed Iterative Reconstruction and Automatic Tube Voltage Modulation (CARE kV) in Abdominal CT
OBJECTIVE To evaluate the feasibility of sinogram-affirmed iterative reconstruction (SAFIRE) and automated kV modulation (CARE kV) in reducing radiation dose without increasing image noise for abdominal CT examination. MATERIALS AND METHODS This retrospective study included 77 patients who received CT imaging with an application of CARE kV with or without SAFIRE and who had comparable previou...
متن کاملLow-Dose High-Pitch CT Angiography of the Supraaortic Arteries Using Sinogram-Affirmed Iterative Reconstruction
OBJECTIVE To prospectively evaluate image quality and radiation dose using a low-dose computed tomography angiography protocol and iterative image reconstruction for high-pitch dual-source CT-angiography (DSCTA) of the supraaortic arteries. MATERIAL AND METHODS DSCTA was performed in 42 patients, using either 120 kVp tube voltage, 120 mAS tube current, 2.4 pitch and filtered back projection, ...
متن کاملImage Quality and Radiation Dose of Lower Extremity CT Angiography Using 70 kVp, High Pitch Acquisition and Sinogram-Affirmed Iterative Reconstruction
OBJECTIVES The purpose of this study was to assess image quality and radiation dose of lower extremity CT angiography (CTA) with 70 kVp, high pitch acquisition and sinogram-affirmed iterative reconstruction (SAFIRE). METHODS Lower extremity CTAs were performed on 44 patients: 22 patients were examined using protocol A (120 kVp, pitch of 0.85 and 120 ml of contrast agent on a first-generation ...