HyFlex nickel-titanium rotary instruments after clinical use: metallurgical properties.
AIM To analyse the type and location of defects in HyFlex CM instruments after clinical use in a graduate endodontic programme and to examine the impact of clinical use on their metallurgical properties. METHODOLOGY A total of 468 HyFlex CM instruments discarded from a graduate endodontic programme were collected after use in three teeth. The incidence and type of instrument defects were anal...
متن کاملMechanical and Metallurgical Properties of Various Nickel-Titanium Rotary Instruments
The aim of this study was to investigate the effect of thermomechanical treatment on mechanical and metallurgical properties of nickel-titanium (NiTi) rotary instruments. Eight kinds of NiTi rotary instruments with sizes of ISO #25 were selected: ProFile, K3, and One Shape for the conventional alloy; ProTaper NEXT, Reciproc, and WaveOne for the M-wire alloy; HyFlex CM for the controlled memory-...
متن کاملMetallurgical characterization of controlled memory wire nickel-titanium rotary instruments.
INTRODUCTION To improve the fracture resistance of nickel-titanium (NiTi) files, manufacturers have introduced new alloys and developed new manufacturing processes for the fabrication of NiTi files. This study aimed to examine the phase transformation behavior and microstructure of NiTi instruments from a novel controlled memory NiTi wire (CM wire). METHODS Instruments of EndoSequence (ES), P...
متن کاملAn in vitro assessment of the physical properties of novel Hyflex nickel-titanium rotary instruments.
AIM To determine several properties including torsional and fatigue limits, as well as torque during canal preparation, of Hyflex, a rotary instrument manufactured from so-called controlled memory nickel-titanium alloy. METHODOLOGY The instruments were tested in vitro using a special torque bench that permits both stationary torque tests according to ISO3630-1 and fatigue limit determination,...
متن کاملDefects in rotary nickel-titanium files after clinical use.
The purpose of this study was to analyze the type and frequency of defects in nickel-titanium rotary endodontic files after routine clinical use, and to draw conclusions regarding the reasons for failure. All of the files (total: 378, Quantec Series 2000) discarded after normal use from a specialist endodontic practice over 6 months were analyzed. Almost 50% of the files showed some visible def...