High bandwidth absorption spectroscopy with a dispersed supercontinuum source.
An optical gas sensor is presented, making use of a dispersed supercontinuum source, capable of acquiring broad bandwidth spectra at ultrahigh wavelength sweep and repetition rates. Wavelength sweeps from 1100 nm to 1700 nm can be performed in 800 ns at a spectral resolution of 40 pm. This is comparable to line-widths of molecular spectra at atmospheric pressure. Quantitative measurements are p...
متن کاملA calibration method for broad-bandwidth cavity enhanced absorption spectroscopy performed with supercontinuum radiation
An efficient calibration method has been developed for broad-bandwidth cavity enhanced absorption spectroscopy. The calibration is performed using phase shift cavity ring-down spectroscopy, which is conveniently implemented through use of an acousto-optic tunable filter (AOTF). The AOTF permits a narrowband portion of the SC spectrum to be scanned over the full high-reflectivity bandwidth of th...
متن کاملCavity enhanced absorption spectroscopy using a broadband prism cavity and a supercontinuum source.
We report the design and construction of a cavity enhanced absorption spectrometer using broadband Brewster's angle prism retroreflectors and a spatially coherent 500 nm to >1.75 microm supercontinuum excitation source. Using prisms made from fused silica an effective cavity reflectivity of >99.99% at 1.064 microm was achieved. A proof of principle experiment was performed by recording the cavi...
متن کاملCavity enhanced absorption spectroscopy of multiple trace gas species using a supercontinuum radiation source.
Supercontinuum radiation sources are attractive for spectroscopic applications owing to their broad wavelength coverage, which enables spectral signatures of multiple species to be detected simultaneously. Here we report the first use of a supercontinuum radiation source for broadband trace gas detection using a cavity enhanced absorption technique. Spectra were recorded at bandwidths of up to ...
متن کاملFourier-transform photocurrent spectroscopy using a supercontinuum light source
Related Articles A capillary absorption spectrometer for stable carbon isotope ratio (13C/12C) analysis in very small samples Rev. Sci. Instrum. 83, 023101 (2012) Standoff detection via single-beam spectral notch filtered pulses Appl. Phys. Lett. 100, 051111 (2012) A two-color tunable infrared/vacuum ultraviolet spectrometer for high-resolution spectroscopy of molecules in molecular beams Rev. ...