Evaluation of the Hitachi 717 analyser
The selective multitest Boehringer Mannheim Hitachi 717 analyser was evaluated according to the guidelines of the Comisión de Instrumentación de la Sociedad Española de Química Clinica and the European Committee for Clinical Laboratory Standards. The evaluation was performed in two steps: examination of the analytical units and evaluation in routine operation.THE EVALUATION OF THE ANALYTICAL UN...
متن کاملEvaluation of a random access analyser: BM/Hitachi 911
The performance of Boehringer Mannheim's BM/Hitachi 911 was evaluated for three months. The mean coeffcient of variation (CV) of the within-run and between-run imprecision of the 16 analytes were less than 1.16% (range 0.47-2.38%) and 1.35% (range 0.62-2.93,%), respectively. A linearity study for the various assays covered clinically important levels. No relevant drift was observed during an ei...
متن کاملEvaluation of a new inhibitor test for isoamylase on Hitachi 705 analyser.
We describe a simple and rapid method for measuring pancreatic and salivary type amylases using the Hitachi 705 analyser. The determination is based on the inhibition of salivary amylase using an inhibitor isolated from wheat germ. The precision of the proposed method was very good: within-day precision varied from 0.4 to 2.5% (CV) and day-to-day precision from 2.2 to 3.7% (CV). The new applica...
متن کاملCharacterization and mathematical correction of hemolysis interference in selected Hitachi 717 assays.
The effect of hemolysis on several assays performed with the Hitachi 717 was quantified by relating the amount of error to the concentration of hemoglobin. Hemolysis interference was judged clinically significant when analyte concentration varied by > 10% from the initial value. Hemolysis interference was significant for alkaline phosphatase, aspartate aminotransferase, alpha-amylase, bilirubin...