Evaluation of immature granulocyte parameters in myeloid neoplasms assayed by Sysmex XN hematology analyzer

نویسندگان

چکیده

Abstract Immature granulocytes (IGs) have significance for the diagnosis of myeloid neoplasms (MNs). The current study aims to use a hematology analyzer evaluate accuracy IG parameters in MNs. Blood specimens from 388 patients with MN, 524 non-hematological (non-HNs), including 109 inflammation and 68 undergoing G-CSF administration, 500 healthy control subjects were analyzed. was assayed by Sysmex XN-9000 (XN) compared manual assessments. A high level agreement between IG% derived XN measurements MN ( r = 0.828, p < 0.0001) revealed but only moderate correlation acute leukemia (AML; 0.597; 0.0001). Bland–Altman bias analysis conducted, results showed that differences considered clinically insignificant. ROC demonstrated good performance IG# (AUC 0.842) 0.885) assessed cut-off values 0.200 × 10 9 /L 1.95%, respectively. are helpful screening MNs even though granulocyte morphological abnormalities may interfere parameter accuracy.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Performance evaluation of the immature granulocyte parameter on the Sysmex XE-2100 automated hematology analyzer.

Recently, St. Anthony's Medical Center Laboratory (SAMC) in St. Louis, Missouri, installed a new automated hematology analyzer, the Sysmex XE-2100. The XE-2100 has an additional differential parameter, the immature granulocyte (IG) parameter. Prior to implementation, the IG parameter was evaluated by comparison to the manual immature granulocyte count. Immature granulocytes counted manually, as...

متن کامل

An evaluation of the Sysmex NE-8000 hematology analyzer.

The Sysmex NE-8000 is a new, fully automated hematology analyzer capable of providing a five-part white blood cell differential count and identifying abnormal specimens. This instrument was evaluated on 5,000 consecutive blood specimens and compared to the Coulter S Plus-IV analyzer and manual differential cell counts to determine the efficacy of its five-cell differential and screening capabil...

متن کامل

Evaluation of Sysmex XN-1000 hematology analyzer for cell count and screening of malignant cells of serous cavity effusion

Over the years, with the advancement in hematology analyzer technology, the use of fluid analysis method has seen a drastic increase in clinical examinations. Cell counting and classification in independent body fluid analysis method are conducted by semiconductor laser flow cytometry and nucleic acid fluorescence staining techniques. This study is to evaluate the efficacy of Sysmex XN-1000 hem...

متن کامل

Performance characteristics of the Sysmex XT-2000i hematology analyzer.

The XT-2000i is a new hematology analyzer combining aperture impedance and optical technologies. It is capable of generating a 30-parameter hematologic assay on peripheral blood at a throughput of 80 samples per hour. We have evaluated the performance characteristics of the XT-2000i analyzer and compared them with those of the state-of-the-art XE-2100 analyzer and the manual differential count....

متن کامل

Evaluation of immature granulocyte counts by the XE-IG master: upgraded software for the XE-2100 automated hematology analyzer.

We evaluated an automated immature granulocyte (IG) count in the peripheral blood with the XE-IG Master (Sysmex Corporation). The XE-IG Master is the upgraded software package for the XE-2100 automated hematology analyzer. Reproducibility tests demonstrated a mean coefficient of variation of 7.02% for the IG percentage (IG%) and 6.93% for the absolute IG count. Correlations of the IG counts wer...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ژورنال

عنوان ژورنال: Journal of Hematopathology

سال: 2022

ISSN: ['1865-5785', '1868-9256']

DOI: https://doi.org/10.1007/s12308-022-00484-w